Recommended separation distances between portable and mobile RF communications equipment and the LMD-
The LMD-2765MD/LMD-2760MD is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the LMD-2765MD/LMD-2760MD can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
LMD-2765MD/LMD-2760MD as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Caution
When you dispose of the unit or accessories, you must
obey the laws in the relative area or country and the
regulations in the relative hospital.
WARNING on power connection
Use a proper power cord for your local power
supply.
1. Use the approved Power Cord (3-core mains
lead) / Appliance Connector / Plug with
earthing-contacts that conforms to the safety
regulations of each country if applicable.
2. Use the Power Cord (3-core mains lead) /
Appliance Connector / Plug conforming to the
proper ratings (Voltage, Ampere). If you have
questions on the use of the above Power Cord
/ Appliance Connector / Plug, please consult a
qualified service personnel.
WARNING on power connection for
medical use
Please use the following power supply cord.
With connectors (plug or female) and cord
types other than those indicated in this table,
use the power supply cord that is approved for
use in your area.
2765MD/LMD-2760MD
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
150 kHz to 80 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
Plug Type
Cord Type
Minimum Rating for Plug
and Appliance Couplers
Safety Approval
*Note: Grounding reliability can only be achieved when the equip-
ment is connected to an equivalent receptacle marked 'Hospital Only'
or 'Hospital Grade'.
For the customers in the U.S.A.
Caution
Federal law (United States of America)
restricts this device to sale by or on the order of
a licensed healthcare practitioner.
m
800 MHz to 2.5 GHz
d = 1.2
d = 2.3
0.12
0.38
1.2
3.8
12
United States and Canada
HOSPITAL GRADE*
Min.Type SJT
Min.18 AWG
10 A/125 V
UL Listed and CSA
0.23
0.73
2.3
7.3
23
7