INDICATIONS
Prosthetic interface with suspension properties for any reason for a major
lower limb amputee; Low to high impact use; community and workplace
ambulatory with ability or potential for cadence variation and ramp/stair
descent. Activity examples: Shopping, longer walks, golfing, recreational
sports.

CONTRAINDICATIONS:

Amputee equipped with a loose fitting socket.
Amputee is unable to follow instructions.
Amputee is unable to ambulate with a lower limb prosthesis.
WARNING!
Make sure that the socket is well fitting before any use; a loose fitting or
leaking socket may never be used with the Unity System.
STRUCTURAL SPECIFICATION
Net weight: approx. 90 g including tube.
Frame construction: Carbon fiber and aluminum frame.
Clearance and dimensions: See IFU for Pro-Flex XC.
FITTING THE UNITY SYSTEM
note: To connect the system to a socket a Unity Socket Valve Kit is
recommended, the use of other applications is the sole responsibility of
the provider.
1. Fit the user with the foot containing the system
2. Cut the tube to correct length and connect to the Unity System and
the socket (using the Unity Socket Valve kit)
3. Use the supplied tube fixture to secure the tube against the foot. The
tube should be located medially rather than laterally to avoid damage
during ambulation.
Once the Unity System has been connected make sure the pump is
working properly; creating vacuum during gait and not leaking air.
DISASSEMBLING / ASSEMBLING THE UNITY SYSTEM
The Unity System can be dissembled in case it needs to be cleaned or for
troubleshooting purposes.
caution: The Unity System should only be disassembled / assembled by
a clinical specialist.
caution: The inlet and outlet (Figure 4) ports on the membrane housing
should not be disassembled unless absolutely necessary. This part of the
system contains miniature duckbill valves that are easily lost and can
cause the Unity System to malfunction if not correctly reassembled.
HOW TO CHANGE MEMBRANE IN THE UNITY SYSTEM
1. Use the Flex-foot shoehorn (or equivalent) to remove the foot from
the foot cover.
2. Disconnect the tube from the Unity Membrane housing.
3. Untighten the screw that fastens the membrane housing to the Pro-
Flex XC pyramid.
4. Follow figure 5.
5. Lubricate the membrane before sliding it into the housing with
Iceross Clean&Simple Spray or equivalent.
HOW TO ASSEMBLE THE UNITY SYSTEM
See figure 6