Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance & Cleaning/Disinfecting; In Case Of Contact With Blood, Saliva, Spatter Or Other Important Soiling - suni SuniCam II Bedienungsanleitung

Inhaltsverzeichnis

Werbung

MAINTENANCE & CLEANING/DISINFECTING
For each new patient, it is essential to use the recommended camera
barrier sheaths. Barrier sheaths can be purchased from Suni.
Before first using the camera, it is imperative to follow the complete
disinfecting procedure.
Any device returning from service or maintenance should follow
the complete disinfecting procedure before being used. Likewise,
any device shipped to Suni for service or maintenance should be
disinfected prior to shipping.
WARNING: Do not directly spray disinfecting products on
the SuniCam II.
Do not use products containing:
• Ammoniac, trichloroethylene
• Dichloroethylene
• Ammonium hydrochloride
• Chlorinated and aromatic hydrocarbon
• Ethylene dichloride
• Methylene chloride
• Ketones
Use of these chemicals can cause deterioration of the plastic
parts.
DISINFECTANT RECOMMENDATIONS:
• Use of surface cleaning and disinfecting wipes.
• We recommend CaviWipes (manufactured by Metrex) or other
cleaning and disinfecting wipes with exact same chemistry.
INSTRUCTIONS:
• Take the cleaning and disinfecting wipe, wring it, and then scrub
the equipment until obtaining visible cleanliness.
• Allow device to dry in the open air.
PRECAUTIONS:
• Do not rinse or immerse device.
• Do not immerse device in a disinfecting liquid.
IN CASE OF CONTACT WITH BLOOD, SALIVA, SPATTER OR OTHER
IMPORTANT SOILING:
• It is strongly recommended to follow a disinfecting process.
• Clean the device with disinfecting wipes.
• Roll up several disinfecting wipes around the handpiece and leave
on for 15 minutes.