Cleaning the reusable jar and lid (PSU and PC)
For optimum user safety, we recommend checking the reusable jar and lid before each use
as natural damage can occur with time. This check consists of a visual check and evacuation
of the jar at maximum vacuum. Faulty jars and lids should be replaced.
Polysulfone (PSU) jar
Disinfect and sterilise for 10 minutes at 134° C / 273° F.
Remove all residues from chemical disinfection agents before autoclaving. Do not stack
when autoclaving. Do not use any phenol-based cleaning agents. With proper use, reusable
jars (PSU) and lids have a service life of 2 years or 200 autoclaving cycles.
After this time, jars and lids must be replaced for safety reasons.
Polycarbonate (PC) jars
Do not autoclave. The maximum disinfection temperature is 100° C / 212° F. Information from
the manufacturer regarding concentration, soaking times or temperature must be observed.
Silicone tubing
Can be autoclaved for 10 minutes at 134° C / 273° F.
Disposal
XV
The Basic 30 comprises metals and plastics and should be disposed of in accordance with
the European directives 2002 / 95 / EC and 2002 / 96 / EC. Additional, local guidelines must also
be observed.
12
EN
Cleaning
Wipe off with a
damp cloth.
These are disposable items and must not be
reused. Replace after every use!
Reprocessing could cause loss of mechanical,
chemical and / or biological characteristics.
Disinfection
With disinfecting
agent (see above)
Sterilisation
Can not be autoclaved