USER MANUAL HYDROGEN LEAK DETECTOR - H2000 PLUS

Sensitivity Settings

GB
Note! Sensitivity Settings only affects
Detection Mode. To calibrate the Analy-
sis Mode, see page 14.
If Direct Sensitivity Adjustment is OFF,
Sensitivity can be adjusted as described
below.
The chosen Sensitivity will only be stored
in memory if adjusted in the menu system.
Select the main menu Sensitivity Settings
as described on page 12.
Sensitivity Settings
1. Press Enter.
Esc
<
>
Enter
2. Select:
Sensitivity,
Auto Range,
1
Direct Sensitivity Adjustment,
Lowest Frequency or
Dead Band using < and >.
Sensitivity Settings
Sensitivity
3. Press Enter.
Esc
<
>
Enter
4. Adjust the desired parameter using the +
and – buttons.
2
3
(5. Press Undo to delete the setting and rev-
ert to the previous value.)
Sensitivity Settings
Sensitivity
6. Press Save to save the set value. The set-
Set Sensitivity
5
ting scale will flash to confirm the setting.
Revert to Detection Mode/Analysis
Mode by pressing Esc twice.
Undo
+
Save
5
4
6