17
Technical Information
Controller Connection
Tab. 4
Pins
Name
15-8
Fault
13-6
Start
12-5
Low Speed
11-4
R1
14-7
R2
218/300
Description
0V = no fail
24V = fail
Configurable output. Default
setting: 24V pump in RAMP
state, 0V = other state.
See win 177 for further
description.
24V = low speed active
0V = low speed not active
Programmable Set Point Out.
See note 2.
Programmable Set Point Out.
See note 2.
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (D)
Electrical
Optically isolated
Vmax (=open) = 24V
Vlow, max = 1,5V
Imax = 60mA
Pin-8 negative
Pin-15 positive
Optically isolated
Vmax (=open) = 24V
Vlow, max = 1,5V
Imax = 60mA
Pin-6 negative
Pin-13 positive
Optically isolated
Vmax (=open) = 24V
Vlow, max = 1,5V
Imax = 60mA
Pin-5 negative
Pin-12 positive
Optically isolated
Vmax (=open) = 24V
Vlow, max = 1,5V
Imax = 60mA
Pin-5 negative
Pin-12 positive
Optically isolated
Vmax (=open) = 24V
Vlow, max = 1,5V
Imax = 60mA
Pin-7 negative
Pin-14 positive
Note
See note 1.
Connect a 10kohm
resistor between the
2 pins before
measuring voltage
A STOP command
(by front Panel or
serial or Start/stop
input) reset the FAIL
status.
See note 1.
See note 1.
See note 1.
See note 1.