Herunterladen Inhalt Inhalt Diese Seite drucken

Chauvin Arnoux C.A 6454 Bedienungsanleitung Seite 71

Schleifenimpedanzmessgerat
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 75
4.5.3
MEASUREMENT PROCEDURE
The measurement is started automatically and is continuous.
4.5.4
MEASUREMENT RESULTS
The measured values and complementary calculated results are described in the table of § 4.1.4 (
position of the switch).
4.5.5
CHARACTERISTICS
4.5.5.1
MEASUREMENT RANGES AND ACCURACY
Particular
reference conditions
Characteristics with an MN 20 current probe:
Display range
Specified measurement
domain
Accuracy
Note: in measurement of I
Characteristics with an C 172 current probe:
Display range
Specified measurement
domain
Accuracy
4.5.5.2
INFLUENCING CONDITIONS
Influencing quantities
Temperature
Relative humidity
Power supply voltage
Frequency (without the
current probe)
Common mode rejection in
50/60Hz AC
:
peak factor = 1.414,
DC component < 0.1%,
operating frequency domain = 15.3 to 450Hz.
400mA
5.0-399.9mA
2%+10pt
, the accuracy is increased by 5%
SEL
400mA
5.0-399.9mA
2%+10pt
Limits of the domain of
use
-10 to + 55 C
10 to 85% RH at 45
6.8 to 10 V
15.3 to 450Hz
0 to 500 V AC
4A
0.400-3.999A
1,5%+2pt
4A
0.400-3.999A
1,5%+2pt
Variation of the measurement
Typical
1%/10 C ± 1pt
2%
1% / V ± 1pt
0.5%
50dB
70
40A
4.00-20.00A
1,2%+2pt
40A
4.00-20.00A
1,2%+2pt
Maximum
2%/10 C + 2pt
3% + 2 pt
2%/ V + 2pt
1%
40dB

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis