Operation

Configure the controller
1. Push Options in the navigation bar to set the general settings, measurements and cleaning
intervals.
2. Select an option.
Option
General settings
Measurement &

Cleaning

Relais & Buzzer
Description
Selects the language. Sets the system time. Shows the system status and contact
information.
Select the box "Measurement" to enable an automatic measurement and set a
measurement interval. Measurement interval: 30 seconds to 1 day (default: 2 minutes)
Select the box "Sensor power down after measurement" to save power.
Push Trigger now to measure a single sample.
Select the box "Cleaning" to enable the cleaning and set the parameters that follow:
• Active: on
• Cleaning interval: 30 seconds to 6 hours (default: 15 minutes)
• Cleaning duration: 5 to 20 seconds (default: 10 seconds)
• Pause before measurement: 5 seconds to 5 minutes (default: 5 seconds)

Measurement

interval
Minimum
30 seconds
Typical
2 minutes
Maximum
1 day
To start a cleaning immediately, push Clean now.
Select the box to enable the relay control and set the parameters that follow:
• Active: on
• Select the parameter
• Action level: Sets the measurement range in which the relay operates.
• Upper level: Sets an alarm to occur when the measurement is higher than the set
point value.
• Lower level: Sets an alarm to occur when the measurement is lower than the set point
value.
The buzzer is a local signal of the instrument. Select the box to enable the instrument
buzzer and set the parameters that follow:
• Active: on
• Select the parameter
• Action level: Sets the measurement range in which the buzzer operates.
• Upper level: Sets the buzzer to occur when the measurement is higher than the set
point value.
• Lower level: Sets the buzzer to occur when the measurement is lower than the set
point value.
Cleaning
Cleaning time
interval
30 seconds
5 seconds
15 minutes
10 seconds
6 hours
20 seconds
Pause before
measurement
5 seconds
5 seconds
5 minutes
English 21