Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
The Blood pressure monitor Medel Control is intended for use in the electromagnetic environment specified below. The cu-
stomer of the user of the Blood pressure monitor Medel Control should assure that it is used in such and environment.
Immunity test
IEC 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Power frequency
3A/m
(50Hz) magnetic
field
IEC 61000-4-8
100029_95142MEDELcontrol_userMANUAL_REV.02_22.03.2017.indd 127
Compliance
Electromagnetic environment – guidance
level
±6 kV contact
Floors should be wood, concrete or ceramic tile. If floor are
±8 kV air
covered with synthetic material, the relative humidity should be
at least 30%.
3A/m
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.