Herunterladen Diese Seite drucken
Vorschau ausblenden Andere Handbücher für CardioServ:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

227 453 01 SA(d/e)-96.03
227 453 01 SA(d/e)-96.03
227 453 01 SA(d/e)-96.03
227 453 01 SA(d/e)-96.03
227 453 01 SA(d/e)-96.03
S e r v i c e a n l e i t u n g
Servicing Instructions
Car
Car
dioSer
dioSer v v v v v
dioSer
Car dioSer
Car
Car
dioSer
Version 3

Werbung

Kapitel

loading

Inhaltszusammenfassung für Marquette CardioServ

  • Seite 1 S e r v i c e a n l e i t u n g Servicing Instructions dioSer dioSer v v v v v dioSer Car dioSer dioSer Version 3 227 453 01 SA(d/e)-96.03 227 453 01 SA(d/e)-96.03 227 453 01 SA(d/e)-96.03 227 453 01 SA(d/e)-96.03 227 453 01 SA(d/e)-96.03...
  • Seite 2 Änderungen und Repara- turen durch MARQUETTE- repairs are carried out by HELLIGE oder durch eine von MARQUETTE-HELLIGE or by persons authorized by MARQUETTE- MARQUETTE-HELLIGE aus- drücklich hierfür ermächtigte Stelle HELLIGE, ausgeführt werden, the electrical installation of the...
  • Seite 3: Inhaltsverzeichnis

    ....Anhang 11 Autohalterung CardioServ 202 307 01 ....Anhang 12...
  • Seite 4: Funktionsbeschreibung

    Elektroden verwendet werden, d.h. außer den herkömmlichen z.B. auch großflachige Klebeelektroden oder Internelektroden. Für den Schrittmacher werden die Klebeelektroden verwendet. Der CardioServ hat für die Start- und Auslösefunktion jeweils zusätzliche Tasten am Gerät. Der Speichervorgang wird mit der Taste Charge an der Apex-Elektrode ausgelöst (bei Intern und Klebeelektroden am Gerät).
  • Seite 5 Patientenleitung erfolgen. Die Betriebsart Sync wird im Display angezeigt, ebenso werden Sync-Marken im Display eingeblendet. Nach der Kardioversion schaltet der CardioServ aus Sicherheitsgründen automatisch auf nichtsynchronisierte Betriebsart um, d.h. vor jeder erneuten Kardioversion muß auf Sync umgeschaltet werden. EKG-Abnahme ist über Paddles oder EKG-Leitung möglich. EKG wird über das LC-Display oder über den eingebauten 50 mm Printer ausgegeben.
  • Seite 6 Funktionskontrolle der Defibrillation durch Entladen in den eingebauten 50-Ohm- Widerstand . Überwachen der Batterieladung bei Netzanschluß durch gelbe LED. Die Monitorfunktion des CardioServ dient zur Registrierung und Überwachung des EKGs und der Sauerstoffsättigung. Die Herzaktionspotentiale werden über eine 5adrige Patientenleitung dem EKG-Vorverstärker zugeführt. Die Auswahl aus 12 Ableitungen ist möglich.
  • Seite 7 Marquette Hellige GmbH CardioServ V. 3 Seite 7 Die Leiterplatten haben folgende Funktionen: Lpl. Stromversorgung (930 117 18) Diese Leiterplatte sorgt für eine 16-V-Versorgung des Gerätes. Hier können kurzzeitig die 10 A für den Ladevorgang bereitgestellt werden. Es ist ein primärgetaktetes Weitbereichsnetzteil.
  • Seite 8 Marquette Hellige GmbH CardioServ V. 3 Seite 8 - Gleichrichten der Hochspannung - Schützen der Hochspannungsdiode - Entladen des Speicherkondensators als Sicherheitsmaßnahme - Messen der Hochspannung am Speicherkondensator - Abgeben der Energie über das Hochspannungsrelais Ein weiterer Block organisiert die Sromversorgung für das gesamte Gerät. Sie wird aus der eingebauten Batterie, aus den Netzteil oder optional von einer extemen Batterie (z.B.
  • Seite 9 Marquette Hellige GmbH CardioServ V. 3 Seite 9 - Erzeugen aller Versorgungsspannungen für den Rekorder - Aufbereiten aller Signale für Printer-Mikrokontroller Weitere Schaltungsteile auf der Lpl. sind : - Erzeugung von 5Volt aus 12Volt über einen integrierten Schaltregler, kurzschlußfest und über Transzorb-Dioden gegen Überspannung gesichert - Schallgeber, Frequenz über SW gesteuert...
  • Seite 10 Marquette Hellige GmbH CardioServ V. 3 Seite 10 Jeder Eingang der Schrittmacherschaltung ist mit 10kW nach +5V abgeschlossen. Dies ist ein Schutz, falls die einspeisenden Schaltungen hochohmige Zustände annehmen können. Dies würde zu unbestimmten Pegeln auf den Eingangsleitungen führen, wodurch die nachfolgenden Bauelemente ein unerwünschtes Verhalten zeigen...
  • Seite 11: Signalnamen

    Marquette Hellige GmbH CardioServ V. 3 Seite 11 2. Signalnamen Schnittstellesignale Lpl. DIGITAL <-> Lpl. ANALOG Stecker AA Anmerkung: I/O aus Prozessorsicht 68332 Signalname Bedeutung Pegel Polarität ENSELD-ENSELA Auswahl der Energiestufe bzw. Geräte-Ein/Aus ---- bzw Stellung EKG DCBA HHHH -> mechanisch gesperrt, Fehlerreaktion HHHL ->...
  • Seite 12 Marquette Hellige GmbH CardioServ V. 3 Seite 12 Signalname Bedeutung Pegel Polarität GEST Zeigt durch L->H-Wechsel an, wenn Generator gestartet high-aktiv wurde und durch H->L-Wechsel wenn gewählte Energie erreicht ist. (Detektion anhand der Schwinger- Rückschlagspannung) ELEC1 - ELEC3 Defi-Elektrodenerkennung ---- HHH ->...
  • Seite 13 Marquette Hellige GmbH CardioServ V. 3 Seite 13 Schnittstellesignale Lpl. DIGITAL <-> Lpl. ANALOG Stecker AB Signalname Bedeutung Pegel Polarität GND_ANA ---- Ground ---- +12V_ANA ---- +12V Versorgung +12V ---- +5V-ANA ---- +5V-Versorgung ---- Schnittstellensignale Lpl. DIGITAL <-> Kamm/Motor-Interfacestecker AC...
  • Seite 14 Marquette Hellige GmbH CardioServ V. 3 Seite 14 Schnittstellensignale Lpl. DIGITAL <-> LCD-Stecker AD (HOSIDEN HLM6323) Anmerkung: I/O aus Grafik-Controllersicht Signalname Bedeutung Pegel Polarität Kontrastspannung var. ---- ---- Negativ Versorung für LCD -23V ---- D3-D0 Serielle Pixeldaten ---- LCD Drive Signal (AC Signal)
  • Seite 15 Marquette Hellige GmbH CardioServ V. 3 Seite 15 Schnittstellensignale Lpl. DIGITAL<->LCD-Stecker AF (OPTREX DMF50174NF- FW) Anmerkung: I/O aus Grafik-Controllersicht Signalname Bedeutung Pegel Polarität Frame Signal high-aktiv ---- Data Latch Signal H->L Clock Signal for Shifting Serial Data H->L Alternate Signal for LCD-Drive...
  • Seite 16 Marquette Hellige GmbH CardioServ V. 3 Seite 16 Schnittstellensignale Lpl. DIGITAL <-> Tastatur-Interfacestecker AH Anmerkung: I/O aus Prozessorsicht 68332 Signalname Bedeutung Pegel Polarit KBROW0_ Tastaturmatrix Spalten low-aktiv KBROW1_ KBROW2_ KBROW3_ KBCOL0_ Tastaturmatrix Zeilen low_aktiv KBCOL1_ KBCOL2_ KBCOL3_ TAENABLE Freigabesignal für Taste Charge/Defib + Defib...
  • Seite 17 Marquette Hellige GmbH CardioServ V. 3 Seite 17 Schnittstellensignale Lpl. DIGITAL <-> RS232-Interfacestecker AI Anmerkung: I/O aus Prozessorsicht 68332 Signalname Bedeutung Pegel Polarität TxDE Transmit Signal (RS232)) +/-12V ---- RxDE_ Recieve Signal (RS232) +/-12V ---- CONNECT1 ---- CONNECT2 Erkennung RS232-Leitung gesteckt...
  • Seite 18 Marquette Hellige GmbH CardioServ V. 3 Seite 18 Receive high-aktiv EKG_TRIG EK-Trigger-Signal high-aktiv RESOXY_ Reset low-aktiv GNDF ---- Ground ---- +5VF ---- +5V Versorgung ---- ---- 12V-Versorgung ---- Schnittstellensignale Lpl. Potentialtrennun g <->Nellcor MP204 Signalname Bedeutung Pegel Polarität Transmit high-aktiv...
  • Seite 19 Marquette Hellige GmbH CardioServ V. 3 Seite 19 3. Blockschema CardioServ Sicherung Netz-Eingang Lpl Display Lpl Netzteil 930 117 17 930 117 18 Digital Analog Defi- Lpl Motor F, G, N Elektroden 388 010 56 303 441 25 EKG- Thermokamm...
  • Seite 20: Montagehinweise

    CardioServ V. 3 Seite 20 4. Montagehinweise Grundsätzlich : Vor dem Öffnen des CardioServ Gerät ausschalten, Netzstecker ziehen, Akku entfernen und ESD-Schutz realisieren. Austauschleiterplatte nur in ESD-Verpackung zurücksenden Öffnen des Gerätes : Die 9 Schrauben auf der Unterseite des Gerätes lösen, Oberschale links neben das Gerät stellen.
  • Seite 21: Fehlersuchhilfe

    Marquette Hellige GmbH CardioServ V. 3 Seite 21 5. Fehlersuchhilfen Gerät läßt sich nicht einschalten bei Netzbetrieb : - 220 Volt Versorgung über Netzteil ok ? Messen an Stecker CA. - Stecker 220Volt Buchse zu Netzteil ok ? - Sicherung in Zuleitung und auf Netzteil ok ? - Leitung Netzteil -- Lpl.
  • Seite 22 Marquette Hellige GmbH CardioServ V. 3 Seite 22 Keine Anzeige auf der LCD - Kontrast schlecht eingestellt ? (Kontrast ändern) - Kontrast-Tasten defekt ? (Tastentest durchführen) - Stecker für Beleuchtung oder Daten ok ? - Beleuchtung defekt ? - Sicherung SI500 (16V) , SI501(Hinterleuchtung) , SI502 (5V) auf Lpl. Digital ok? - Display defekt ? - Lpl.
  • Seite 23 Marquette Hellige GmbH CardioServ V. 3 Seite 23 Fehlersuchhilfe Lpl Schrittmacher Auf der Lpl Schrittmacher kann nichts abgeglichen werden. Beim Einbau der Platte ist auf richtiges Stecken der Leitungen zu achten; besonders bei der roten Schrittmacherleitung vom Stecker C ist die richtige Polung auf Stecker L und M wichtig.
  • Seite 24: Abgleichsanweisung

    Marquette Hellige GmbH CardioServ V. 3 Seite 24 6. Abgleichanweisung MP1/ MP2/ MP3/ R 523 R 520 R 521 R 522 SI502 SI502 SI500 Lpl Digital 388 028 09 R PCB Digital 388 028 09 R...
  • Seite 25 Marquette Hellige GmbH CardioServ V. 3 Seite 25 Motor-Abgleich Was abgleichen Womit messen? Wie Testmode Woran drehen? Wie groß und Was sonst kontrollieren? einstellen? genau? beachten? Motordrehzahl Mit Stroboskop 1)Konfiguration Geschwindigkeits- Stroboskop- Motor soll bei kleiner Last (LED) aufrufen (Taste...
  • Seite 26 Marquette Hellige GmbH CardioServ V. 3 Seite 26 Steht kein Stroboskop zur Verfügung: Was abgleichen Womit messen? Wie Testmode Woran drehen? Wie groß und Was sonst kontrollieren? einstellen? genau? beachten? Motordrehzahl Mit Linial das 1)Konfiguration Geschwindigkeits- < 1% Motor soll...
  • Seite 27 Marquette Hellige GmbH CardioServ V. 3 Seite 27 Kontrolle der Versorgungsspannungen auf Lpl. Digital Was abgleichen Womit messen? Wo messen? Wie groß und Was sonst kontrollieren? genau? beachten? 5V-Versorgung Spannungsmeß- zwischen +5V 5,0V +/-2% gerät. und GND +22V- Spannugsmeß- zwischen +22V...
  • Seite 28 Marquette Hellige GmbH CardioServ V. 3 Seite 28 R 507 SI 501 R 501 Pin 9 kodiert Lpl Analog 388 028 10 R PCB Analog 388 028 10 R...
  • Seite 29 Marquette Hellige GmbH CardioServ V. 3 Seite 29 Abgleich Energie Was abgleichen Womit messen? Wo messen ? Woran drehen? Wie groß und Was sonst kontrollieren? genau? beachten? In 5o Ohm ab- Defit-Tester R509 .100J +-2J R508 nach links gegebene auf Minimum...
  • Seite 30 Marquette Hellige GmbH CardioServ V. 3 Seite 30 Abgleich EK-Verstärker Was abgleichen Womit messen? Wo messen ? Woran drehen? Wie groß und Was sonst kontrollieren? genau? beachten? Offset EK- Maßstab Display R502 Versatz der Verstärkung Verstärker Kabel optisch Nullinie < 1 mm zwischen 0.5...
  • Seite 31: Leiterplattenverwendung

    Marquette Hellige GmbH CardioServ V. 3 Seite 31 7. Leiterplattenverwendung in verschiedenen CardioServ-Varianten CardioServ CardioServ CardioServ CardioServ CardioServ 101 117 01 101 117 05 101 117 03 101 117 04 101 117 01 Version 2.0/2.1 Version 3.0/3.01 Version 3.0/3.01 Version 3.0/3.01 Version 3.0/3.01...
  • Seite 32: Ersatzteilliste

    101 117 80 AT CardioServ SCP 910 227 432 05 Serviceanleitung CardioServ SCP 910 Defi-Elektrode/n 217 304 03 Defibrillationselektrode, extern, mit Schreiberstart, für CardioServ Registrierteil 384 015 81 AT Registrierteil kompl. für CardioServ 303 439 76 Thermoschreibkamm 303 434 78...
  • Seite 33 Marquette Hellige GmbH CardioServ V. 3 Seite 33 Gehäuse 303 442 85 Gehäuseschale oben 303 443 61 Geäuseschale oben ( Esaote ) 430 517 93 Klebeschild, Bedienhinweis, englisch 430 517 94 Klebeschild, Bedienhinweis, französisch 430 518 09 Klebeschild, Bedienhinweis, italienisch...
  • Seite 34 227 432 01 Gebrauchsanweisung CardioServ (d) 223 287 01 Patientenleitg.,3adr., 2,2 m, o.HF Schutz, Klammern, IE (4.G.u.VICOM) 226 130 02 Registrierpapier,T,55mm,Rolle 40 m für CardioServ 931 098 68 Zubehörtasche CardioServ 931 098 88 Gerätetasche für CardioServ, Nylongewebe, orange, mit Trageriemen...
  • Seite 35 930 117 36 AT Lpl. Spo2 388 030 88 Lpl. Potentialtrennung SPO2 303 443 58 Adapterleitung für SpO2-Sensoren (NELLCOR) an CardioServ, 2,9 m lang 303 443 57 SPO2 Eingangsbuchse komplett verdrahtet für Cardio Serv Folientastatur 390 001 07 Folientastatur ohne Schrittmacher...
  • Seite 36 Marquette Hellige GmbH CardioServ V. 3 Seite 36 9. Technische Beschreibung Betriebsarten Dauer der Ladung des Speicherkonden- sators bei Energieeinstellung 360 J: unsynchronisiert (Defibrillation zu belie- – bei Netzbetrieb typisch 8 s bigem Zeitpunkt) – bei voll geladener Batterie typisch 8 s, synchronisiert (Kardioversion) –...
  • Seite 37 Marquette Hellige GmbH CardioServ V. 3 Seite 37 Sicherheitsentladung Synchronisation Entladung des Kondensators über internen mit dem beliebig gepolten EKG-Signal des Lastwiderstand: Patienten: bei Nichtauslösen des Defibrillationsim- Mindestamplitude des EKG-Signals für pulses innerhalb 30 s nach der Aufla- sichere Triggerung etwa 0,50 mV bei 80...
  • Seite 38 Marquette Hellige GmbH CardioServ V. 3 Seite 38 EKG-Signaleingang über Patientenlei- Erkennung von Paceimpulsen tung – Impulsdauer d > ca. 0,1 ms. <2.0 ms – Pace-Markierung vorzeichenunabhän- Über EKG-Elektroden, automatische Um- schaltung auf EKG-Elektroden bei ange- – Impulsamplitude a ± 20 ... ± 700 mV schlossener Patientenleitung;...
  • Seite 39 Marquette Hellige GmbH CardioServ V. 3 Seite 39 Signalübertragung Alarm-Management Signaleingang –> Verstärkung –> Signalab- elektronische Alarmauslösung tastung –> Analog-Digital-Umsetzung –> Digitalverarbeitung –> Bildschirm und Recor- bei Überschreitung der eingestellten Grenzwerte (mindestens 4 s lang): aku- stisches Signal ertönt (abschaltbar), Empfindlichkeit umschaltbar: 1/2 - 1 - 2 Hinweis „HF Alarm“, Alarmregistrierung...
  • Seite 40 Marquette Hellige GmbH CardioServ V. 3 Seite 40 – Herzfrequenz Rollendurchmesser max. 60 mm (Rolle ® – Ableitung mit 40 m HELLIGE CONTRAST – Filter – Datum Auflösung Schreibkamm vertikal 6 – Uhrzeit Punkte/mm – Papiergeschwindigkeit horizontal 24 Punkte/mm –...
  • Seite 41 Marquette Hellige GmbH CardioServ V. 3 Seite 41 Schrittmacher Betriebsbereitschaft Betriebsarten: Demand, Fix, Over- 4 s incl. automatischem Selbsttest pace Gebrauchslage Impuls-Frequenz: 30…150 P/min (Overpace bis 300 P/ beliebig Umgebungsbedingungen Impuls-Strom: 0…200 mA (bei 500 Ω), bis zu einer Betrieb...
  • Seite 42: Servicing Instructions

    Servicing Instructions CardioServ Version 3 227 453 01 SA(e)-96.03...
  • Seite 43 Caution: During repairs/service interventions, observe the protective measures against damage due to ESD. MARQUETTE HELLIGE considers itself responsible for the effects on safety, reliability, and per-formance of the equip- ment, only if: assembly operations, extensions, readjustments, modifications, or repairs are carried out by...
  • Seite 44 Wiring Diagram 101 117 03 / 04 / 05 ....Appendix 11 Car-Holder for CardioServ 202 307 01 ....Appendix 12...
  • Seite 45: Functional Description

    Adhesive electrodes are used for the pacemaker. The CardioServ also has Start and Trigger buttons on the instrument. The charge operation is actuated with the Charge key on the apex electrode (for internal and adhesive electrodes on the instrument).
  • Seite 46 The operating mode Sync appears on the display and Sync marks are superimposed on the display. Following cardioversion, for safety reasons, the CardioServ switches automatically over to the unsynchronized operating mode, i.e., before each cardioversion the instrument must be switched over to Sync.
  • Seite 47 50-ohm resistor. Monitoring battery charge via the connector for power cord by the yellow LED. The monitoring function of the CardioServ records and monitors the ECG and oxygen saturation level. The cardiac-action potentials are fed to the ECG preamplifier via a 5-wire patient cable. Selection can be made from 12 leads. The ECG curve is displayed on the LC screen with an illuminated background.
  • Seite 48: Pcb. Analog 388 028 10

    Marquette Hellige GmbH CardioServ V. 3 Page 7 The PCBs have the following functions: PCB Power Supply (930 117 18) This PCB supplies the 16-V instrument power supply. 10 A can be supplied for short-term capacitor charging. It is a primary clock-rated wide-range power supply system.
  • Seite 49 Marquette Hellige GmbH CardioServ V. 3 Page 8 - rectification of the high voltage - protection of the high-voltage diode - discharging the storage capacitor as a safety precaution - measurement of the high voltage on the storage capacitor - energy delivery via the high-voltage relay A further unit organizes the entire instrument´s power supply.
  • Seite 50 Marquette Hellige GmbH CardioServ V. 3 Page 9 - generation of power supply voltages for the recorder - processing of all signals for the printer microcontroller Further circuit components on the PCB are : - generator of 5 V from 12 V via an integrated switching controller, short-circuit-...
  • Seite 51 Marquette Hellige GmbH CardioServ V. 3 Page 10 Every pacemaker circuit input is equipped with a 10k Ω resistor after +5V supply. This is a protective measure if the feed-in circuits can adopt high-impedance states. This would lead to uncertain levels on the input cables, which could result in the...
  • Seite 52 Marquette Hellige GmbH CardioServ V. 3 Page 11 2. Signal names Interface Signals PCB DIGITAL <-> PCB ANALOG Connector AA Note: I/O from processor 68332 viewpoint Signal Name Meaning Level Polarity ENSELD-ENSELA Selection of energy level or device On/Off ----...
  • Seite 53 Marquette Hellige GmbH CardioServ V. 3 Page 12 Signal Name Meaning Level Polarity GEST With a L->H change shows when the generator was high-active started and with H->L change when the selected energy level is reached. (Detection by oscillator reverse voltage)
  • Seite 54 Marquette Hellige GmbH CardioServ V. 3 Page 13 Interface Signals PCB DIGITAL <-> PCB ANALOG Connector AB Signal Name Meaning Level Polarity GND_ANA ---- Ground ---- +12V_ANA ---- +12V Power Supply +12V ---- +5V-ANA ---- +5V Power Supply ---- Interface Signals PCB DIGITAL <-> Printhead/Motor Interface Connector AC...
  • Seite 55 Marquette Hellige GmbH CardioServ V. 3 Page 14 Interface Signals PCB DIGITAL <-> LCD Connector AD (HOSIDEN HLM6323) Note: I/O from Graphics Controller viewpoint Signal Name Meaning Level Polarity Contrast Voltage var. ---- ---- Negative Power Supply for LCD -23V...
  • Seite 56 Marquette Hellige GmbH CardioServ V. 3 Page 15 Interface Signals PCB DIGITAL<->LCD Connector AF (OPTREX DMF50174NF- FW) Note: I/O from Graphics Controller viewpoint Signal Name Meaning Level Polarity Frame Signal high-active ---- Data Latch Signal H->L Clock Signal for Shifting Serial Ddta H->L...
  • Seite 57 Marquette Hellige GmbH CardioServ V. 3 Page 16 Interface Signals PCB DIGITAL <-> Keypad Interface Connector AH Note: I/O from processor 68332 viewpoint Signal Name Meaning Level Polarity KBROW0_ Keypad Matrix Columns low-active KBROW1_ KBROW2_ KBROW3_ KBCOL0_ Keypad Matrix Lines...
  • Seite 58 Marquette Hellige GmbH CardioServ V. 3 Page 17 Interface Signals PCB DIGITAL <-> RS-232 Interface Connector AI Note: I/O from processor 68332 viewpoint Signal Name Meaning Level Polarity TxDE Transmit Signal (RS232)) +/-12V ---- RxDE_ Receive Signal (RS232) +/-12V ----...
  • Seite 59 Marquette Hellige GmbH CardioServ V. 3 Page 18 Interface PCB Pacemaker<->PCB Potential Separation Signal Name Meaning Level Polarity Transmit high-active Receive high-active EKG_TRIG ECG Trigger Signal high-active RESOXY_ Reset low-active GNDF ---- Ground ---- +5VF ---- +5V Supply ---- ----...
  • Seite 60 Marquette Hellige GmbH CardioServ V. 3 Page 19 3. Block Diagram CardioServ Fuse Power Input PCB Power PCB Display Supply 930 117 17 930 117 18 Digital Analog Defi PCB Motor F, G, N Electrodes 388 010 56 303 441 25...
  • Seite 61 CardioServ V. 3 Page 20 4. Assembly Instructions General Information: Before opening the CardioServ switch the instrument off, disconnect power cord plug, remove battery and implement ESD protection measures. Return replacement PCB in ESD packaging only Opening the instrument : Undo the 9 screws on the bottom of the device, place upper case shell to the left next to the instrument.
  • Seite 62: Troubleshooting Hints

    Marquette Hellige GmbH CardioServ V. 3 Page 21 5. Troubleshooting Hints Device cannot be switched on during line-power operation: - 220-V power supply via power supply unit OK ? Measure at connector CA. - Connector 220-V plug to power supply unit OK ?
  • Seite 63 Marquette Hellige GmbH CardioServ V. 3 Page 22 No display on LCD - Contrast poorly adjusted ? (change contrast) - Contrast keys defective ? (perform key test) - Connector for illumination or data OK ? - Illumination defective ? - Fuse SI500 (16V) , SI501(background illumination) , SI502 (5V) on PCB Digital...
  • Seite 64 Marquette Hellige GmbH CardioServ V. 3 Page 23 Troubleshooting hints for PCB Pacemaker No adjustments can be made on the PCB Pacemaker. When inserting this PCB make sure it is plugged in correctly; especially in the case of the red pacemaker cable from connector C it is important to observe the correct polarity on connectors L and M.
  • Seite 65 Marquette Hellige GmbH CardioServ V. 3 Page 24 6. Adjustment Instructions MP1/ MP2/ MP3/ R 523 R 520 R 521 R 522 SI502 SI502 SI500 Lpl Digital 388 028 09 R PCB Digital 388 028 09 R...
  • Seite 66 Marquette Hellige GmbH CardioServ V. 3 Page 25 Motor Adjustment What to adjust What to How to activate Where to turn? How much and What else to or to check? measure with? testing mode? exact? note? Adjust motor Flash at motor...
  • Seite 67 Marquette Hellige GmbH CardioServ V. 3 Page 26 No Stroboscope Available : What to adjust What to How to activate Where to turn? How much and What else to or to check? measure with? testing mode? exact? note? Adjust motor...
  • Seite 68 Marquette Hellige GmbH CardioServ V. 3 Page 27 Control of the Power Supply Voltages on PCB Digital What to adjust What to Where to How much and What else to or to check? measure with? measure? exact? note? 5-V power Voltmeter.
  • Seite 69 Marquette Hellige GmbH CardioServ V. 3 Page 28 R 507 SI 501 R 501 Pin 9 kodiert Lpl Analog 388 028 10 R PCB Analog 388 028 10 R...
  • Seite 70 Marquette Hellige GmbH CardioServ V. 3 Page 29 Energy Adjustment What to adjust What to Where to Where to turn? How much and What else to or to check? measure with? measure? exact? note? Energy Defi tester R509 .100J +-2J...
  • Seite 71 Marquette Hellige GmbH CardioServ V. 3 Page 30 EC Amplifier Adjustment What to adjust What to Where to Where to turn? How much and What else to or to check? measure with? measure? exact? note? Offset EC Optical scale Display...
  • Seite 72 Marquette Hellige GmbH CardioServ V. 3 Page 31 7. PCB Application in Different CardioServ Versions CardioServ CardioServ CardioServ CardioServ CardioServ 101 117 01 101 117 05 101 117 03 101 117 04 101 117 01 Version 2.0/2.1 Version 3.x Version 3.x...
  • Seite 73: Pcb. Pacemaker Partly Assembled

    Marquette Hellige GmbH CardioServ V. 3 Page 31 Pcb. usage on different CardioServ CardioServ CardioServ CardioServ CardioServ CardioServ CardioServ 101 117 01 101 117 05 101 117 03 101 117 04 101 117 01 101 117 09 USA Version Version 2.0/2.1...
  • Seite 74 ESAOTE CardioServ V. 3 Page 32 Pcb. usage on different CardioServ Version CardioServ CardioServ CardioServ CardioServ CardioServ Esaote Esaote Esaote Esaote Esaote 101 117 02 101 117 08 101 117 06 101 117 07 101 117 02 Version 2.0 / 2.1 Version 3.x...
  • Seite 75: Spare Parts List

    User Manual Version 3 ( german ) 227 446 04 User Manual Version 3 ( USA ) Defi paddles 217 304 03 Defibrillation Electrode, external, with recorder start, for CardioServ 217 308 01 Electrode Lead f.Defi-El.“internal“ without contact insert, for SCP 913 926 074 97 Handle...
  • Seite 76 Marquette Hellige GmbH CardioServ V. 3 Page 34 Housing 303 442 85 Upper shell 303 443 61 Upper shell ( Esaote ) Version 303 444 46 Upper shell USA Version 430 517 93 Sticker “operation information“, english 430 517 94 Sticker „operation information“, french...
  • Seite 77 226 130 02 Recording Chart,T,55mm, roll 40 m for CardioServ 931 098 68 Accessory Bag for CardioServ 931 098 88 Instrument Bag for CardioServ, nylon fabric, orange, with belt Software 443 295 02 EPROM Cardioserv V 2.1 443 295 05 EPROM Cardioserv V 3.02...
  • Seite 78 930 117 36 Exchange pcb.SPO2 388 030 88 Pcb. Potential Separation for SPO2 303 443 58 Adapter Cable for SpO2 Sensors (NELLCOR) to CardioServ + EAGLE10xx 303 443 57 SPO2 Inputconnector kompletely wired for Cardio Serv Keypad 390 001 07...
  • Seite 79: Technical Specifications

    Marquette Hellige GmbH CardioServ V. 3 Page 37 9. Technical Specifications Operating modes capacitor charging time for energy setting of 360 J: – from power line: typically 8 s non-synchronized (defibrillation at any time) – from fully charged battery: typically 8 s synchronized (cardioversion) –...
  • Seite 80 Marquette Hellige GmbH CardioServ V. 3 Page 38 Safety discharge ECG signal input via paddles capacitor discharge via internal load resistance: ECG signal acquisition via defibrillator paddles, ECG trace displayed on LCD, automatic switching to ECG when the defibrillation shock is not triggered electrodes when patient cable is connected;...
  • Seite 81: Signal Transmission

    Marquette Hellige GmbH CardioServ V. 3 Page 39 differential DC voltage compatibility ±1 V resolution 320 x 240 pixels (pitch of .36 x .36 mm) common-mode rejection (CMRR) R,L referred to N 65 dB, N referred to chassis >110 dB displayed image can be rotated 180 °...
  • Seite 82: Alarm System

    Marquette Hellige GmbH CardioServ V. 3 Page 40 Alarm system electronic release of alarm After the ECG recording, a patient ID sheet is printed indicating name, date of birth, user, comments, date, when the HR violates one of the set limits for at time, energy and alarm limits.
  • Seite 83: Operating Position

    Marquette Hellige GmbH CardioServ V. 3 Page 41 Power saturation: 0 to 100 %, in 1-% increments from the power line frequency: 0 to 250 BPM, in 1-BPM increments 95 V to 240 V, 49 Hz to 65 Hz alarm limit: off, 15 to 100 %...
  • Seite 84 Marquette Hellige GmbH CardioServ V. 3 Page 42 Environment Operation under the following conditions regarded as normal: temperature between 0 and +40 °C rel. humidity between 30 and 95%, no conden- sation atmospheric pressure between 700 and 1060 hPa Storage and transport temperature between -20 and +60 °C...
  • Seite 85: Car-Holder For Cardioserv 202 307 01

    388 030 88 SpO Complete Wiring Diagram 101 117 01...32 Wiring Diagram 101 117 01...13 Master Record Index 101 117 .. Installation Instructions for the CardioServ Wall Mount System 202 307 02 Car-Holder for CardioServ 202 307 01 504 658 18...
  • Seite 86 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 87 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 88 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 89 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 90 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 91 388 028 09 P Bl. 0 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67136 Antrag vom Index Datum/Name LPL. DIGITAL Datum/Name gez. 22.03.95/MLM gepr. PCB DIGITAL Entw.
  • Seite 94 388 031 06 P Bl. 0 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67136 Antrag vom Index Datum/Name LPL. DIGITAL Datum/Name gez. 27.03.95/MLM gepr. PCB DIGITAL Entw.
  • Seite 96 388 031 05 P Bl.0 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67136 Antrag vom Index Datum/Name LPL. DIGITAL Datum/Name gez. 22.03.95/MLM gepr. 23.03.95/MLM PCB DIGITAL Entw.
  • Seite 100 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 101 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 102 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 103 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 104 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 105 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 106 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 109 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 110 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 111 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 112 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 113 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 114 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 115 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 117 388 031 14 P Bl. 0 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67150 Antrag vom Index Datum/Name LPL. SCHRITTMACHER Datum/Name gez. 16.01.96/JOK gepr. PCB PACEMAKER Entw.
  • Seite 120 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 121 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 122 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 123 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 124 388 028 74 P Bl.0 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67150 Aend. -Nr. Index Datum/Name LPL. SCHRITTMACHER Datum/Name gez. 16.01.96/JOK gepr. 16.01.96/HWM PCB PACEMAKER Entw.
  • Seite 127 388 030 88 P HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67177 Antrag vom Index Datum/Name LPL. POTENTIALTRENNUNG SPO2 Datum/Name gez. 5.12.94/HWM gepr. PCB POTENTIAL SEPARATION SPO2 Entw.
  • Seite 129 388 030 88 P HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN F67177 Antrag vom Index Datum/Name LPL. POTENTIALTRENNUNG SPO2 Datum/Name gez. 5.12.94/HWM gepr. PCB POTENTIAL SEPARATION SPO2 Entw.
  • Seite 135 HELLIGE GMBH Postfach 728 D-79007 Freiburg AENDERUNGEN Antrag vom Index Datum/Name Datum/Name gez. gepr. Entw.
  • Seite 136 HELLIGE GMBH Postfach 728 390 001 45 P D-79007 Freiburg AENDERUNGEN F67136 Antrag vom Index Datum/Name FOLIENTASTERTUR Datum/Name gez. 04.04.96/MLM gepr. KEYPAD Entw.
  • Seite 141 Installation Instructions for the CardioServ Wall Mount System 202 307 02 * The mounting system has 4 bore holes both at * Please be sure to use the red feet (angular the top and bottom edges, allowing it to be groove) at the bottom, i.e., on the side with the...
  • Seite 145 European Headquarters World Headquarters Asia Pacific Marquette Hellige GmbH Marquette Medical Systems, Inc. Marquette Medical Systems Postfach 728 8200 West Tower Avenue 26/F, Catic Plaza 79007 Freiburg • Germany Milwaukee, WI 53223 • USA 8 Causeway Road, Hong Kong Tel.