Table of contents
Safety information
Refer to the installation user manual for general safety information, hazard descriptions and
precautionary labels descriptions.
Show the active errors, warnings and reminders
The display background and the status indicator light will change to red when an error occurs or to
yellow when a warning occurs.
• Error-red display background and status indicator light. A significant problem occurred that
affects the instrument operation. The current measurement stops and the analyzer goes into
shutdown mode.
• Warning-yellow display background and status indicator light. An event occurred that can cause
a future problem. The analyzer continues to operate.
• Reminders-wrench symbol shows on the display and yellow status indicator light. The time for a
maintenance task has passed.
Do the steps that follow to show the active errors, warnings and reminders:
Note: To see all of the possible errors, warnings and reminders and their help information, push diag, then select
ANALYZER HELP.
1. Push diag, then select DIAGNOSTICS. The active errors, warnings and reminders are shown.
2. Select an error, warning or reminder from the list.
3. Select an option.
Option
ACKNOWLEDGE Acknowledges the selected error, warning or reminder. When all of the active errors,
VIEW HELP
Maintenance
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Description
warnings and reminders are acknowledged, the display background changes back to
white and the status indicator light changes back to green.
Shows help information for the selected error, warning or reminder.
D A N G E R