Figure 8 Probe Connector Wiring - Fluke 914X-Serie Bedienungsanleitung

Field metrology well
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 82
Probe Connector
1
6
2
3
RTD Sensor
Figure 8
Probe connector wiring
A two-wire probe can also be used with the reference thermometer. It is connected by attach-
ing one wire to both pins 1 and 2 of the plug and the other wire to both pins 4 and 5. If a shield
wire is present, it should be connected to pin 3. Accuracy may be significantly degraded using a
two-wire connection because of lead resistance.
4-20mA Connectors (2)
The 4-20mA connectors allow current and/or voltage probes to be connected for measurement
of associated devices.
PRT/RTD Connector (3)
The 4-wire PRT/RTD connectors allow the user to connect 3-wire and 2-wire (with jumpers,
see Figure 9 on next page) PRT/RTDs to the readout. The correct wiring for the 4-wire PRT/
RTD is shown on the instrument. Figure 9 shows the correct wiring for a 2 or 3-wire PRT/RTD
(see Section 1.5.2 Immunity Testing on page 8 for information on using clamp-on ferrites).
5
4
Shield