Mechanical installation
M
A
Nominal (rated) force in kN
0.5...10
20
50
100
200
Tab. 6.2
Mounting dimensions when using one knuckle eye
6.3.3
Assembly without adapter
The U2B is delivered with an adapter that can be removed, if the sensor is to
be bolted to a construction element by the thread in the measuring body.
The construction element must have the following properties:
S Flatness: 0.01 mm
S Maximum roughness: Ra =0.8 μm
20
ØG
H8
S
f8
K*
84 ... 86.4
131.6 ... 133
158.2 ... 160.8
244 ... 246
270.2 ... 272.4
A00382_03_YI0_06 HBM: public
Mounting base,
customer side
*
Recommended dimension, taking
minimum thread reach into consideration
U2B