4 Starting the milling process
8
NOTICE
Selecting the milling instruments
Following installation, the inserted combination of milling instruments
may still be unknown to the software. In this case, a dialog box will
automatically open where you then must select the milling instruments
currently inserted in the milling unit:
The dialog box is open.
➢ Select the milling instrument which you have inserted in the left
motor mount from the
➢ Select the milling instrument which you have inserted in the right
motor mount from the
1. Select the required material from the
then appears.
2. Select the required block size.
NOTICE
Error during touch process
Incorrect specification of the block manufacturer or block selection may
lead to failure of the touch process.
3. Confirm your choice by clicking the
The milling unit then moves to the insertion position.
4. Depending on the configuration, you will now be prompted to enter
the bar code (see also "Entering the bar code [ → 10]").
5. Press the catch of the milling chamber door and open the door.
NOTICE
Error message during touch process!
Always be sure to insert the ceramic block that you selected in the
"Select block"
dialog box. Otherwise an error message will be displayed
during the touch process.
6. Place the selected ceramic block in the block clamp.
7. Clamp the ceramic block with the ball pressure screw. Also use the
block clamp tool (see "Using a manual block clamp").
8. Close the milling chamber door and confirm the procedure by clicking
"Start"
.
The estimated time required for the milling process will then
appear in a message window.
NOTICE
Aborting the milling process
You can abort the milling process at any time by pressing the
button.
9. When the milling process has been completed, open the milling
chamber door.
10. Remove the restoration.
Sirona Dental Systems GmbH
Supplement to Operating Instructions
"Left"
list.
"Right"
list.
"Select block"
"OK"
button.
D 3493.201.05.03.02 08.2011
dialog box which
"Stop"
63 22 346 D 3493