Modul.: enables or disables modulation. The input is done
with the arrow keys left and right.
Mod.-factor: modulation factor between 2 and 254. This
command is available only when modulation is enabled.
The (multiplication) product of frequency and modulation
factor must not exceed 2.500.000 Hz. E.g. at a frequen-
cy of 100.000 Hz, a 25-fold modulation is just permissible,
a frequency of 500.000 Hz may be modulated only up to
5-fold, a frequency of 9842 Hz can be modulated by 254,
thus approaching the modulation threshold (2.499.868
Hz).
A new number is entered using the numeric keys. The pre-
vious number is automatically deleted. If only individual di-
gits are to be deleted, this is possible by pressing the arrow
key to the left.
Impul.: enables or disables impulse discharge. The input
is done with the arrow keys left and right. If impulse is
activated, the following commands are available:
:
Active time
time (in seconds) during which the output fre-
quency is active. The input is done with the arrow keys left
and right.
Pause time: time (in seconds) when the frequency output
is inactive. The input is done with the arrow keys left and
right.
During the defining process the program can be executed
by pressing the START-button.
Now this program is defined completely.
If you press the EXIT button to leave the editing menu, the
device asks whether you want to save the program with
OK or to discard with EXIT.