Wobble Time
Duration of a complete wobble cycle (fre-
quency 1 → frequency 2 → frequency 1). This input is only
available if Wobble is enabled. The input is done with the
arrow keys left and right. The time must be between 6 and
30 seconds.
Cons. amps: activation of constant current control
This input is not available when Const amps has been dis-
abled in the Settings.
In the constant current control the device maintains the
current flowing through the body constant. For this purpo-
se, the voltage is increased or decreased according to the
instantaneous resistance of the skin and the electrodes.
The input is done with the arrow keys left and right.
Intensity: Intensity in milliamps (1/1000 amps) or volts, de-
pending on whether the constant current control is enabled
or not. The input is done with the arrow keys left and right.
Duration: Execution time of the program in minutes and
seconds. Using the arrow key to the left the current time
can be deleted. A new time can be entered by using the
numeric keys. The comma key is used to separate mi-
nutes and seconds. If the comma key is not pressed, the
number entered will be used in seconds.
For Example: The input of 5,15 results in a time of 5 minu-
tes and 15 seconds.