Herunterladen Inhalt Inhalt Diese Seite drucken

Device Care; Cleaning The Cuff; Accuracy Test; Disposal - Microlife BP A150 AFIB Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für BP A150 AFIB:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 34
- contamination and dust
- direct sunlight
- heat and cold
• The cuffs are sensitive and must be handled with care.
• Do not exchange or use any other kind of cuff or cuff connector
for measuring with this device.
• Only pump up the cuff once fitted.
• The function of this device may be compromised when used
close to strong electromagnetic fields such as mobile phones or
radio installations and we recommend a distance of at least
1 m. In cases where you suspect this to be unavoidable, please
verify if the device is working properly before use.
• Do not use this device if you think it is damaged or notice
anything unusual.
• Never open this device.
• If the device is not going to be used for a prolonged period the
batteries should be removed.
• Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

Carefully remove spots on the cuff with a damp cloth and soapsuds.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!

Accuracy test

We recommend this device is tested for accuracy every 2 years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.

12. Guarantee

This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
BP A150 AFIB
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
• Batteries and parts that become worn with use are not included.
• Opening or altering the device invalidates the guarantee.
• The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
• The cuff has a functional guarantee (bladder tightness) for 2 years.
Please contact your local Microlife-Service (see foreword).

13. Technical Specifications

Operating conditions: 10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions: -20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
393 g (including batteries)
Weight:
152 x 92 x 42 mm
Dimensions:
Measuring procedure: oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
0 - 299 mmHg
range:
1 mmHg
Resolution:
pressure within ± 3 mmHg
Static accuracy:
± 5 % of the readout value
Pulse accuracy:
Voltage source:
4 x 1.5 V alkaline batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
approx. 920 measurements
Battery lifetime:
(using new batteries)
IP20
IP Class:
EN 1060-1 /-3 /-4; IEC 60601-1;
Reference to
IEC 60601-1-2 (EMC); IEC 60601-1-11
standards:
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
7
EN

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis