Capacitance transmitters
Vorschau ausblenden Andere Handbücher für SITRANS LC300:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
Mounting Location
Recommendations
• Easy access for programming the unit using the rotary switch and buttons.
• An environment suitable to the housing rating and materials of construction.
Precautions
• Keep unit out of path of falling material. Protect probe from falling material by installing a
shield.
• Consider material surface configuration when installing unit.
• Tensile load must not exceed probe or vessel rating. On cable units, do not exceed the
tensile strength of the cable at 1900 kg / 4188 lbs.
Buildup of material or condensation in the active shield area does not affect
Note:
operation.
Wall Restriction
50 mm
(2") min
Sensors must be at least 50 mm (2") from the wall or tank bottom and must be 500 mm (20") apart.
.
Page EN-4
SITRANS LC300 – QUICK START MANUAL
Multiple Units
50 mm
(2") min
500 mm
(20") min
500 mm (20") min