Herunterladen Inhalt Inhalt Diese Seite drucken

Connection Of Water; Connection To The Drain - Getinge 607 Montage-Und Installationsanleitung

Flusher disinfector
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Connection of water

1. Flush through the water pipes that are to be
connected to the machine, to prevent blockage of the
filter and valves.
2. Connect the wash disinfector to the cold and hot
water.
The water connections must satisfy the following
requirements:
607/608
Kallvatten
Anslutning
Tryck
Flöde
Varmvatten
Temperatur
Anslutning
Tryck
Flöde
617/618
Kallvatten
Anslutning
Tryck
Flöde
Varmvatten
Temperatur
Anslutning
Tryck
Flöde
3. Seal the connections with Teflon tape.

Connection to the drain

Connect the disinfector to the drain. The drain can be
connected either rearwards or downwards. The pipe
diameter is 110 mm.
6
Installation
15 (1/2") m m
70-800 kPa
20 l/min
45-60 °C
15 (1/2") m m
70-800 kPa
20 l/min
20 (3/4") mm
250-800 kPa
35 l/min
45-60 °C
15 (1/2") m m
250-800 kPa
30 l/min
Edition 040210
Shut-off valve
Filter
V553
V661
Drain connection P in wall
Drain connection S in floor
V662
5013246-00