balance will return to normal weighing mode.

LIMITS INDICATORS

• When the weight exceeds the high limit, the indicator "over" will light on.
• When the weight is between the low and high limits, the indicator "accept"
will light on.
• When the weight is below the low limit, the indicator "low" will light on.
Note: If the display is instable, the limits indicators won´t light on.
Attention: When the low limit is greater than the high limit, the display will
show the message Err2, in this case, please cancel the operation.

CALIBRATION

Resolution selection
1. There is a switch on the bottom of the scale protected by a cover and two
screws. Move the swich towards the keyboard.
2. Turn on the scale pressing the key
3. Keep pressed the key
4. Press the key
5. Press the key
6. Press the key
7. Press the key
14
until the display shows the message "rAngE"
until you see "d"
to access.
repeatedly to see the resolution values available.
to select and confirm.
.