6 Starting Up

6.2 Installation

6.2.1 Fix and connect the pyrometer

DANGER:
There is danger of injury and possible equipment damage by connection of cables under power!
Never connect cables under power! Make sure that the power supply is switched off before connection
of the cables to the pyrometer!
Install the pyrometer as follows:
1. The selected controller at the ordering is already correctly build in and adjusted.
2. Double check stable set up of the panel instrument case.
3. Double check stable set up of the fibre optic.
4. Connect the fibre optic with the optical head and with the panel instrument case. (Pay attention to the
color bend radius)
5. Connect the connecting cable according to the cable assignment with the used devices.
6. Connect the power cable.
The pyrometer is ready for use and the power supply can be switched on.

6.2.2 Align pyrometer

The pyrometer is equipped with an LED pilot light for accurate alignment of the sensor with the object to be
measured. Align the pyrometer with the object to be measured as follows:
1. Switch on the supply voltage to the pyrometer.
2. Allow a starting time of approx. 5 minutes for thermo-stabilization. Than the pyrometer has stabilized and
is ready to work with the given accuracy.
3. The pilot light is always on. Make sure that there is nothing in the path of rays.
Note:
In order to avoid measuring errors, the area of the pilot light must not be larger than the object to be
measured.
4. If you have a pyrometer with vario optic head, adjust the necessary measuring distance (see Setting the
optic (section 7.1) at page 13).
Note:
The fixed optics cannot be adjusted.
The pyrometer is thus aligned and ready for the temperature measurement.