Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Specifications - Nutricia Flocare Infinity PLUS Gebrauchsanleitung

Vorschau ausblenden Andere Handbücher für Flocare Infinity PLUS:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19

TECHNICAL SPECIFICATIONS

• This Flocare
®
Infinity
ned to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emis-
sions and immunity and are in compliance with Directive 93/42/EEC.
0344
• IEC 601: Class II
• Microprocessor controlled
• Type BF Shock Protection (EN60601-1) degree of protection against electrical shock. No electrical connection
to the patient. Drop from any height of 90cm shall not damage pump operation.
• Charger:
input 100-240V AC / 50-60 Hz / 0,4A Max.
output 5 V DC 2,4 A
• Battery: internal rechargeable Lithium ion battery, 3,6 V DC 2000 mAh
• Battery capacity: 24 hours at 125 ml/h
• IPX 5: EN60529 degree of protection against water entering the enclosure. Water jets from any direction
shall have no effect on the pump.
• Dimensions: 140 x 95 x 35 mm
• Weight: approx. 392 g
• Accuracy flow rate: ±5,0% with appropriate Flocare
• Humidity:
• Temperature:
• Atmospheric pressure:
• Occlusion detection pressure: Upstream occlusion:
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated,
varies from 0.5 to 1.5 ml which relates to an air bubble with an approximate length in the tubing of
9 - 26 cm.
• The Flocare
Infinity
®
• The use of other accessories, adaptors and cables than listed within this manual may result in increased
emissions or decreased immunity of the equipment of the Flocare
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in the accompanying documents.
• Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equip-
ment. If the Flocare
be observed to verify normal operation.
• In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number.
• The pump and its electrical accessories (AC Adapter Charger and Data Download Accessory Cable),
should be disposed of according to local standards and regulations governing the disposal of electronic
waste (e-waste). Other accessories can be disposed of or recycled as standard non-hazardous waste.
ZMNL900617_AW.indd 90
+ pump in combination with the adapter and the Flocare
Medical Device Directive: Class IIa
Operation mode:
Storage:
Operation mode:
Storage and transportation: -20°C to +65°C
Operation mode:
Storage and transportation: 50-106 kPa
Downstream occlusion:
+ pump can safely be operated on commercial aircraft.
®
Infinity
+ pump is used adjacent to or stacked with other equipment, the pump should
®
®
Infinity
pump set
30% to 75% noncondensing
10% to 95 % noncondensing
+5°C to +40°C
70-106 kPa
-34 kPa (tolerance 21 kPa)
83 kPa (tolerane 21 kPa)
Infinity
+ pump.
®
-90-
UK-IE
Infinity
Nurse Call are desig-
10/11/16 16:18

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis