• As the battery level decreases the pump may not be able to deliver tube feeds at rates greater than 300 ml/h,
even though some battery capacity remains. The pump will alarm and indicate "BATT". If this happens, either
recharge the battery or select a lower rate to complete the feed cycle.
• In case of battery failure, caused by, for example, an excessive temperature condition (which may resolve
itself with time) or by a complete battery failure, the E, F and battery symbol will flash. Turn the pump "off"
and contact your healthcare provider or refer the pump to service.
The power connector at the side of the pump offers the option of connecting the pump to an external alarm
system or Patient Data Monitoring System (PDMS).
• The Flocare
Infinity
®
This external alarm system may be with open or closed contact. Follow the instructions for use of the
Flocare
Infinity
Nurse Call to connect the pump.
®
• The Flocare
Infinity
®
system. Follow the instructions for use of the Flocare
• Detergent and disinfectant cleaning agents and wipes can damage plastic surfaces of medical devices
when they are not compatible with the surface material. Damaged surfaces may compromise the ability
to decontaminate medical devices adequately and / or may interfere with device function and affect their
integrity and performance.
• Avoid harsh or agressive cleaners/ disinfectants such as acetone, benzene or iodine.
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm
soapy water, a 5% bleach solution in water, a multipurpose disinfectant cleaner or a soft cloth.
• The Flocare
Infinity
®
merge the pump!
• Always maintain the rollers on the rotor in a clean state to ensure they spin freely.
• The adaptor normally does not require cleaning. When desired, a dry or slightly damp cloth may be used
to clean the outside surface of the adaptor. Make sure the adaptor is disconnected from the wall outlet.
• Pump disposal: At the end of their service lifetime , the pump and its electrical accessories
(AC Adapter Charger and Data Download Accessory Cable), should be disposed of accor-
ding to local standards and regulations governing the disposal of electronic waste (e-waste).
Other accessories can be disposed of or recycled as standard non-hazardous waste.
• The manufacturer recommends an inspection of the pump at an authorised service centre every 2 years.
• If any fault occurs during use, or if the pump is dropped, it should be checked by authorised technical per-
sonnel prior to use.
• Replace the Flocare
growth of harmful bacteria. Dispose of Flocare
• Do not use the Infinity alarm systems to trigger actions related to secondary (electrical) medical devices
(e.g. a volumetric or syringe pump).
• Do not use pump functions (e.g. the fill set function) for any other purpose than described in this manual,
ZMNL900617_AW.indd 85

CONNECTION TO AN EXTERNAL ALARM SYSTEM

Nurse Call (corporate code 35752) enables connection to an external alarm system.
PDMS Cable (corporate code 35776) enables connection to an external PDMS

MAINTENANCE AND CLEANING

+ pump may be rinsed by holding under a stream of warm, clean water. Do not sub-

WARNINGS AND CAUTIONS

Infinity
pump set every 24 hours to maintain delivery accuracy and prevent the
®
®
®
Infinity
PDMS Cable to connect the pump.
Infinity
disposable sets properly, as required by local law.