• Press the "start/stop" key to pause the pump.
• Change the program by pressing the required key (ml/h, volume ml or bolus ml/int h) and adjust using
the "+" or "-" keys.
• Check all installed parameters to assure correct programming of the pump.
• Restart the pump by pressing the "start/stop" key again.
All parameters and values can be cleared one after the other in the following way:
• Press "start/stop" to pause the pump.
• Select the parameter that needs to be cleared.
• Press the "clr" key to clear the memory. The parameter or value returns to its default value:
ml/h
= 0 ml/h
volume ml = CONT = No dose set. Pump will feed continuously until feed container is
bolus ml
= NO
int h
= NO
info
= 0 ml
In order to keep a clear overview of the daily amount of feed delivered, clear the total volume delivered as
each daily feeding schedule is started, as follows:
start/stop
clr
another patient, all settings (rate and dose) can be returned to default and the volume delivered can be
set at zero by simply pressing the "clr" key and keeping it pressed for 5 seconds.
• The Flocare
Infinity
®
• The battery symbol in the display is automatically replaced by the plug symbol when the adapter is
plugged in. The "fuel gauge" then indicates that the battery is charging by displaying the segments in
an ascending low to high pattern starting with the leftmost segment.
This pattern continuously repeats while the pump is charging.
• To check the status of the battery, disconnect the charger from the pump and turn the
pump on. The bars between E and F (E=Empty, F=Full) represent the "fuel gauge" of the
battery. Each bar is approximately 1/4 of a full battery charge. If 2 bars appear the
battery is half full and remaining operating time is approximately 12 hours at a flow rate
of 125 ml/h.
• In case of power failure the pump automatically switches to battery powered operation.
• When during functioning the last block is gone, the battery symbol will blink to indicate that there is approxima-
tely 1 hour of charge left. The display will flash "BATT" every 3 seconds, alternating with the active display and
the pump will beep every 2 seconds to remind the user of the low battery charge condition. Plug in the charger
to continue to run and recharge the battery.
ZMNL900617_AW.indd 84
TO CHANGE THE FEEDING PROGRAM DURING FUNCTIONING

CLEARING THE MEMORY

empty or pump is stopped.
= No bolus volume set. The pump will run in continuous feeding mode.
= No bolus interval set.
= Volume delivered is cleared.
• Switch on the pump.
• Immediately after the self-test the pump displays total volume delivered.
• Press the "clr" key. The total volume delivered returns to "0 ml".
• The pump is in the "hold" mode ready for programming.
• When another feeding program needs to be started, or the pump will be used for

BATTERY OPERATION

+ pump is equipped with a Lithium-ion battery.
-84-
UK-IE
10/11/16 16:18