as this may cause the pump to incorrectly calibrate.
• No modification of this equipment is allowed.
• The AIR alarm should not be used to indicate an end of dose. NOTE: do not mute the audio alarm for end
of dose alarm in these instances ('end of dose' will appear in the pump screen).
• Non-liquid food ingredients and mixed or reconstituted foods may have a tendency to occlude flow within
the feeding tube and feeding set. They may also have the tendency to adhere to the tubing wall and
obstruct the enteral feeding pump optical air-in-line sensor. Clinicians and caregivers should carefully
monitor the pump during use with these type of feedings.

ALARM FUNCTIONS AND SAFETY FEATURES

In case any of the problems listed on the problem solver chart occur, the pump delivers an audible and visual
alarm and stops working. The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "start/stop" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "start/stop".
The problem solver chart gives a clear explanation of the alarms.
Using methods other than those described may cause the pump to function incorrectly.
Condition
NO SET
• The set is not fitted or wrongly fitted
• The pressure sensor area is dirty (
PUSH STRT
• The pump has been untouched in
END OF DOSE
• The pump administered the installed
PROG
• No flow rate is installed.
Flow rate = 0 ml/h
• The installed parameters in a bolus pro-
gram cannot form a correct solution.
DOOR
• The door is not correctly closed.
AIR
• The pump detected air in the pump
ZMNL900617_AW.indd 86

PROBLEM SOLVER CHART

Always follow the instructions below in case a problem occurs.
Cause
in the pump.
Fig 1, position 5 & 6)
hold mode for 3 minutes or more.
dose = volume.
set.
Correction
• Stop the alarm by pressing "start/stop".
• Insert the Flocare® Infinity™ feeding set into the pump as
indicated on the blister packaging (or refer to Fig 5a - 5d)
and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
• Stop the alarm and prolong the hold mode with another 3
minutes by pressing "start/stop" key.
• Program the pump and start it by pressing the "start/
stop".
• The programmed dose/volume has been administered
• Press start/stop to switch off the alarm
• Press INFO and then CLR to reset the administered
volume (to reset the flow rate and dose as well, press the
CLR button for 3 seconds)
• Make sure the correct flow rate is programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Make sure there is no excessive foam in the reservoir and
ensure there is enough feed in the reservoir for the set
therapy.
• If the feeding reservoir is empty, replace the reservoir and
continue feeding. If necessary, prime the set (FILL SET).