Limitations of the alternative status page:
• Set and actual value of power are not displayed and the set value is only indirectly accessible
• Set and actual value of resistance are not displayed and the set value is only indirectly accessible
• No access to the settings overview (MENU button) while the DC input is on
In alternative status page mode, the set values of power and resistance are not adjustable while
the DC input is switched on. They can only be accessed and adjusted in SETTINGS while the
DC input is off.
Rules for manual handling of the HMI in alternative status page mode:
• The two rotary knobs are assigned to voltage (left knob) and current (right knob) all the time, except for menus
• Set values input is the same as in standard status page mode, with knobs or by direct input
• Regulation modes CP and CR are displayed alternatively to CC at the same position
3.4.8

The meter bars

Additionally to the actual values being displayed as numbers, meter bars for U, I and P can be enabled in the
MENU. Refer to „3.4.3.8. Menu "HMI settings"" to see where to enable the meter bars in the MENU.
Depiction:
Standard status page with meter bar
3.4.9

Switching the DC input on or off

The DC input of the device can be manually or remotely switched on and off. This can be restricted in manual
operation by the control panel being locked.
Switching the DC input on during manual operation or digital remote control can be disabled by
pin REM-SB of the built-in analog interface. For more information refer to 3.4.3.1 and example
a) in 3.4.3.7. In such a situation, the device will show a notification in the display.
► How to manually switch the DC input on or off
1.
As long as the control panel (HMI) is not fully locked press the button ON/OFF. Otherwise you are asked to
disable the HMI lock (simple unlock or by entering the PIN, if activated in menu "HMI Lock").
2.
The ON/OFF button toggles between on and off, as long as a change is not restricted by any alarm or the
device being in "Remote".
► How to remotely switch the DC input on or off via the analog interface
1.
See section "„3.5.4 Remote control via the analog interface (AI)" on page 58.
► How to remotely switch the DC input on or off via the digital interface
1.
See the external documentation "Programming Guide ModBus & SCPI" if you are using custom software, or
refer to the external documentation of LabView VIs or other documentation provided by EA Elektro-Automatik.