Power
frequency
(50/60 HZ)
magnetic
field IEC
61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
Guidance and Declaration - Electromagnetic Immunity
The Blood pressure monitor is intended for use in the specified
electromagnetic environment. The customer or the user of the Blood pressure
monitor should assure that it is used in such an environment as described
below.
IEC6060
Immunit
1
y test
level
3
150 kHz
Conduce
to
d RF
80 MHz
IEC61000
outside
-4-6
ISM
bands
Radiated
3 V/m 80
RF
MHz to
IEC61000
2.5 GHz
-4-3
3 A/m
Complianc
test
e level
Vrms
3 Vrms 150
kHz to
80 MHz
outside
ISM bands
3 V/m 80
MHz to
2.5 GHz
in UT)
for 5 s
3 A/m
Electromagnetic
guidance
Portable
communications equipment should
be used no closer to any part of the
system, including cables, than the
recommended separation distance
calculated
appropriate for the frequency of the
transmitter.
separation
=
d
1
.
2
Recommended
distances:
80 MHz~800 MHz:
800MHz-2.5GHz:
Where, P is the maximum output
power rating of the transmitter in
watts
(W)
14
Power
frequency
magnetic fields should
be
at
characteristic
typical location in a
typical commercial or
hospital environment.
environment
and
mobile
from
the
equation
Recommended
distances:
P
separation
=
d
1
.
2
=
d
2
.
3
according
to
levels
of
a
-
RF
P
P
the

Werbung

Inhaltsverzeichnis
loading