Herunterladen Inhalt Inhalt Diese Seite drucken

Leica FS C Bedienungsanleitung Seite 15

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Oblique Incident Illumination
• Screw the special articulated arms onto the
left and right of the macroscope carrier (5.1).
• Screw the light guide with ø 4.5 mm (Fig. 6)
onto the cold light illuminator (8.1).
• Place the cold light source next to the stand
or onto the shelf that is to be screwed onto
the base of the comparison macroscope (7.1).
• Insert both light guide ends into the left and
right receptacles of the special articulated
arms (5.2) and tighten them using an Allen
wrench.
• Mount and fasten the focusing optics (6.1).
Fig. 5
Special articulated arms for oblique incident illumi-
nation
1
Fastening screws for the articulated arm
2
Receptacle for light guide
1
2
Fig. 6
Light guide ø 4.5 mm
1
Focusing optics
Fig. 7
Shelf for the cold light illuminator
1
Fastening screws for the shelf
Fig. 8
Cold light illuminator KL2500 LCD
1
Connection for light guide
7. Assembly
1
1
1