elbow and container when installing/re-installing.
Do not reverse direction of filter. Additional filters
(7305D-608 12/pack) may be purchased from your
authorized DeVilbiss Healthcare provider.
NOTE– Do not substitute any other material for this
bacteria filter. Substitution may lead to contamination
or poor performance; use only DeVilbiss filters.
NOTE– Bacteria filter must be changed between
patients.

CLEANING INSTRUCTIONS

Preparation
1. Shut off unit using power switch and allow vacuum
to drop. Disconnect power source from the DC
input receptacle on the unit.
2. Disconnect tubing and remove container from
holder.
3. Carefully remove lid and empty contents.
NOTE-Container should be emptied and cleaned after
each use.
WARNING
To prevent possible risk of infection from
contaminated cleaning/disinfection solutions,
always prepare fresh solution for each cleaning
cycle and discard solution after each use.
Disposable Collection Container with
Internal Filter Cartridge
The disposable collection container and lid are meant
for single-patient use.
1. Remove filter cartridge and 4-3/8" tubing and set
aside.
NOTE– Filter MUST NOT get wet. The filter material
cannot be removed from the elbow (figure A).
A
WARNING
Do not remove float ball from lid. If removed, float
ball may pose a choking hazard (figure B).
B
C L E A N I N G
2. Wash container, lid and splash guard in warm
water/dishwashing solution. Rinse with clean,
warm water.
3. Soak in 1 part vinegar (>=5% acetic acid
concentration) to 3 parts water (131˚F-149ºF/55˚C-
65˚C) solution for 60 minutes. Rinse with clean,
warm water and air dry.
NOTE– The disassembled container may also be
washed in a dishwasher, top shelf only, using a cycle
with a water temperature between 131ºF-149ºF/55ºC-
65ºC.
Disposable Collection Container with
External Bacteria Filter
1. Remove bacteria filter, 3-3/8" tubing, and
connection elbow and set aside.
2. Wash jar and lid in a solution of warm water with a
mild, liquid detergent (e.g. Dawn or Palmolive) and
rinse with clean, warm tap water.
3. Soak in 1 part vinegar (>=5% acetic acid
concentration) to 3 parts water (131˚F-149ºF/55˚C-
65˚C) solution for 60 minutes. Rinse with clean,
warm water and air dry.
NOTE– The disposable collection container and lid are
meant for single-patient use.
NOTE– The disassembled container may also be
washed in a dishwasher, top shelf only, using a cycle
with a water temperature between 131ºF-149ºF/55ºC-
65ºC.
Reusable Collection Container with
External Bacteria Filter
1. Remove bacteria filter, 4-3/8" tubing, and
connection elbow and set aside. Remove o-ring
and overflow valve from lid.
2. Wash jar, lid, o-ring, and overflow valve in a
solution of warm water with a mild, liquid detergent
(e.g. Dawn or Palmolive) and rinse with clean,
warm tap water.
3. For single patient use: After washing, disinfect
using one of the following methods:
a. Soak in a solution of 1 part vinegar (>=5%
acetic acid concentration) to 3 parts water
(131˚F-149ºF/55˚C-65˚C) for 60 minutes. Rinse
with clean, warm water and air dry in a clean
environment.
b. Wash with rubbing alcohol and air dry in a
clean environment.
c. Wash with a commercial (bacterial-germicidal)
disinfectant. Follow disinfectant manufacturer's
recommended dilution rates and instructions
carefully.
4. For multi-patient use: After washing:
a. After parts are completely dry, place jar and lid
in autoclave with the open end down. Ensure
parts are not touching. Run one sterilization
5
LT-1819