Herunterladen Inhalt Inhalt Diese Seite drucken

wtw Cond 330i Bedienungsanleitung Seite 8

Leitfähigkeits-taschenmeßgerät
Vorschau ausblenden Andere Handbücher für Cond 330i:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Determining/Setting the cell constants
C
A L
1/
cm
0 7
4
3
Tref25
Cal
C L
E
L
6
8 0
S/
1/
0 1
6
4
Tref25
ARng
nLF
C L
E
L
9
2 5
S/
1/
0 1
0
0
Tref25
ARng
nLF
8
l Press <RUN/ENTER>. The AutoRead measurement begins.
l If the measured value is stable, the instrument displays the determined
cell constants and the calibration evaluation. The measuring instrument
automatically stores the cell constants.
l Switch to the measuring mode with <M>.
Procedure for manually setting the cell constant:
l Press the <CAL> key repeatedly until the
l Then press the <RUN/ENTER> key.
l Press the <CAL> key until a measured value and a value for the cell con-
stant appears in the required setting range.
cm
l Set the value of the cell constant with < > < >.
cm
l Switch to the measuring mode with <M>.
Procedure for selecting fixed cell constants 0.010 cm
l Press the <CAL> key repeatedly until the
l Then press the <RUN/ENTER> key.
l Press the <CAL> key until a measured value and the cell constant
-1
0.010 cm
cm
l Switch to the measuring mode with <M>.
cm
appear.
Cond 330i/340i
CELL display appears.
-1
:
CELL display appears.

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cond 340i

Inhaltsverzeichnis