Error codes and countermeasures
Error codes are indicated by a single letter in the lower right corner of the display. However, these
only occur in the menus M00, M01, M02, M03, M90 and M08. The following chart shows the error
codes and countermeasures.
Error
code
R
I
J
H
Q
F
1
2
3
K
Message in window
M08
System Normal
Detect No Signal
Hardware Error
PoorSig Detected
Frequ OutputOver
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
Adjusting Gain
Empty pipe
© PCE Instruments
Reason
No error
(1) No signal
(2) Sensors installed
improperly
(3) Too much fouling
(4) Liner too thick
(5) Sensor cable not
Properly
connected
Hardwareproblem
(1) Schlechtes Signal
(2) Sensoren falsch
angebracht
(3) Zu viel Bewuchs,
zu starke
Verschmutzung
(4) Auskleidung vom
Rohr zu dick
(5) Sensorkabel nicht
korrekt
angeschlossen
Die
Ausgangsfrequenz
liegt außerhalb des
erlaubten Bereiches
(1) Vorübergehende
Probleme mit dem
RAM oder RTC
(2) Permanente
Probleme mit der
Hardware
Das Gerät stellt
gerade die
Signalverstärkung
(Gain) neu ein; die
Zahl gibt den
aktuellen
Arbeitsfortschritt an
(1) Keine Flüssigkeit
in der Rohrleitung
(2) Einstellungsfehler
im Menü M29
37
Countermeasure
- - -
(1) Change
measuring
location
(2) Clean measuring
location
(3) Check the cables
Contact PCE
Instruments
(1) Change
measuring
location
(2) Clean measuring
location
(3) Check the cables
(4) Check the
contact gel
Check the values in
the windows M67,
M68 and M69. Enter
higher values in
window M69.
(1) Restart the device
(2) Contact PCE
Instruments
- - -
(1) Choose a pipe
that contains
liquid
(2) Enter "0" in
window M29.