KEMA 02 ATEX 1006 / XA174F-D
Safety instructions:
Installation
10
Explosion hazardous area
Zone 1
Electronic
insert
Electronic
insert
Tank
Hazardous
area
Zone 0
Local equipotential bonding
Fig. 1
Certified associated apparatus
Category
Type of protection
Max. working pressure
Process temperature
Housing
F12
Option
Remote display
e.g. FHX40
Service interface
• Install the device according to the manufacturer's instructions and any other valid standards and regulations.
• The intrinsically safe input power circuit of the device is isolated from ground potential and has a dielectric
strength of at least 500 Vrms with respect to it.
• The relationship between the permitted ambient temperature for the electronics housing, dependent on the
range of application and the temperature classes is shown in the tables (Tab. 1).
• Continuous duty temperature of the cable ≥ Ta +5 K.
• The type of protection changes as follows when the devices are connected to certified intrinsically safe
circuits of Category Ex ib for Equipment Groups IIC and IIB: Ex ib IIC or Ex ib IIB.
• The pertinent guidelines must be observed when intrinsically safe circuits are connected together acc.
IEC/EN 60079-14 (Proof of Intrinsic Safety). (E.g. when using Commubox or handheld terminal DXR375
or other certified apparatus).
FMU42, FMU44
• Avoid electrostatic charging of the sensor (e.g. do not rub dry and install outside the filling flow).
Option: (
certified
)
- Remote display FHX40
- Service interface
Tank
Hazardous
area
Zone 1
Uo = 30 V
Io = 300 mA
Po = 1 W
II 1/2 G
or II 2 G
Ex ia IIC T4...T6
FMU40, FMU41: max 0.3 MPa
FMU42, FMU44: max. 0.25 MPa
max. 80 °C
–40 °C ≤ Tu ≤ +80 °C
KEMA 02 ATEX 1203
Commubox
with associated ToF cable
Prosonic M
Non-hazardous area
Certified associated apparatus
Certified associated apparatus
[Ex ia] IIC
Sensor in Zone 0, housing in Zone 1
or sensor and housing in Zone 1
KEMA 02 ATEX 1006
Optionally with or without VU331 display
and operating module
Observe associated Safety Instructions
(XA...)
Observe associated Safety Instructions
(XA...)
Endress+Hauser
XA174en01