Note: All voltage and current levels set via the keyboard are displayed on a separate 0·3" 4-digit
display. This entry preview system ensures that the user can observe the value entered before it
is effected thus avoiding possible error. The display is also used for setting additional functions.
When the output switch is on and no other function is selected, the display shows output power in
Watts.
DIGITAL INTERFACES
Operational Functions:
RS232:
GPIB:
OUTPUT SPECIFICATIONS - REMOTE OPERATION
Output Voltage Setting:
Output Current Setting:
Setting Accuracy:
Output Switch:
Readback Resolution:
Readback Accuracy:
RESPONSE TIME OVER RS232/GPIB:
Interface:
Power Supply:
GENERAL
AC Input Voltage:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Size:
Weight:
Safety & EMC:
Set voltage, set current, set OVP; set output on/off; read output
voltage; read output current; read output power.
Variable baud rate, 9600 baud maximum, 9 pin D-connector (female).
Fully compatible with ARC (Addressable RS232 Chain) system.
Conforming with IEEE488.1 and IEEE488.2.
12 bit resolution (10mV steps).
12 bit resolution (10mA steps).
Voltage: ±(0·1% + 10mV); Current: ± (0.2% + 20mA)
Electronic by interface command.
Voltage: 10mV over the entire range.
Current: 10mA over the entire range.
Voltage: ± (0.2% of reading + 1 digit);
Current: ± (0.5% of reading + 1 digit).
<15ms (single command, buffer empty).
An internal time constant, T, (typically 22ms) governs the settling time
of a step voltage increase. Settling time to within 1% of the step
change = 4.6T, to 0.1% = 6.9T, to 0.01% = 9.2T; for example, after a
10V step the output will be within 1 digit (10mV = 0.1%) of its new
value in typically 150ms. For load current of 1 Amp or more, settling
times for downward steps will be very similar; however, response
times will be longer at low loads.
110V-120V AC or 220V-240V AC ±10%, 50/60Hz.
Installation Category II.
600VA max.
5°C to +40°C, 20% to 80% RH.
–40°C to +70°C.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
200 x 140 x 385mm (WxHxD); half rack width x 3U height;
(optional rack mounting kit available)
5.5kg.
Complies with EN61010-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this
instrument via http://www.aimtti.com/support (serial no. needed).