DECVV
Sequential command
Operation complete message generated immediately after execution.
Decrement the output voltage by the delta voltage value and verify that the voltage is within ±3
counts or 5% of the target value. If the value of the output voltage is outside the range for the
specified output the value is set to the minimum value allowed, no error is generated.
DECV
Sequential command
Operation complete message generated immediately after execution.
This command is identical to DECVV above except that no verification of the output voltage is
performed, thus saving the possible 500ms required to read back the output voltage. This is
useful when it is known that the slew time will be short or during constant current operation.
INCI
Sequential command
Operation complete message generated immediately after execution.
Increment the set current by the delta current value. If the value of the set current is outside the
range for the specified output the value is set to the maximum value allowed, no error is
generated.
DECI
Sequential command
Operation complete message generated immediately after execution.
Decrement the set current by the delta current value. If the value of the set current is outside the
range for the specified output the value is set to the minimum value allowed, no error is
generated.
POWER
Sequential command
Operation complete message generated immediately after <rmt> is sent.
Returns the power being output from the instrument, in watts, in <nr2> numeric format. The
syntax of the response is
Example:
OP <nrf>
Sequential command
Operation complete message generated immediately after execution.
Sets the output status ON or OFF. If the value of <nrf>, after rounding, is 0 the output status will
be set to OFF if 1 the output status will be ON. If the value of <nrf> is not 0 or 1 execution error
119 (Value out of range) will be generated.
DAMPING <nrf>
Sequential command
Operation complete message generated immediately after execution.
Sets the current meter damping to OFF or ON. If the value of <nrf>, after rounding, is 0 the meter
damping will be set to OFF, if 1 the meter damping will be set to ON. If the value of <nrf> is not 0
or 1 execution error 119 (Value out of range) will be generated.
If the output power is 175.3 watts the response to the command POWER? will be
175.3W<rmt>.
<nr2>W<rmt>
27