Verfügbare Sprachen

∗WAI
Sequential command.
Operation complete message generated immediately after execution.
Wait for operation complete true. As all commands are completely executed before the next is
started this command takes no additional action.
Instrument Specific Commands
The commands in this section are additional to those specified by IEEE Std. 488.2 as Common
Commands.
VV <nrf>
Sequential command
Operation complete message is generated once the new output voltage has settled to within ± 3
digits or 5% of the required value. If the voltage fails to settle within 5 seconds the Operation
Time-out Error (bit 3 in the Standard Event Status Register) will be set and the operation
complete message will then be generated.
Set the voltage to <nrf> and verify that the voltage is within ± 3 counts or 5% of the target value.
The value of <nrf> must be in Volts; no multipliers are allowed. If the value of <nrf>, after
rounding, is outside the range of the specified output an execution error will be generated and
the corresponding error number will be placed in the Execution Error Register, 100 (Maximum set
voltage value exceeded) if the value is too large and 102 (Minimum set voltage exceeded) if the
value is too small.
V <nrf>
Sequential command.
Operation complete message generated immediately after execution.
This command is identical to VV <nrf> above except that no verification of the output voltage is
performed, thus saving the possible 500ms required to read back the output voltage. This is
useful when it is known that the slew time will be short or during constant current operation.
I <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Set current limit to <nrf>. The value of <nrf> must be in Amps; no multipliers are allowed. If the
value of <nrf>, after rounding, is outside the range of the specified output an execution error will
be generated and the corresponding error number will be placed in the Execution Error Register,
101 (Maximum set Amps value exceeded) if the value is too large or 103 (Minimum set Amps
value exceeded) if the value is too small.
OVP <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Set OVP to <nrf>. The value of <nrf> must be in Volts; no multipliers are allowed. If the value of
<nrf>, after rounding, is outside the range of the specified output an execution error will be
generated and the corresponding error number will be placed in the Execution Error Register,
108 (maximum set OVP value exceeded) if the value is too large or 107 (Minimum set OVP value
exceeded) if the value is too small.