Herunterladen Inhalt Inhalt Diese Seite drucken

Kommunika Ons-Protokoll Rs232/C - LR-Cal FLUID 100 Anleitung

Portables temperaturkalibrierbad
Inhaltsverzeichnis

Werbung

Portable Kalibrierbäder
MANUAL
FLUID 100 und FLUID 100-N

11. Kommunika ons-Protokoll RS232/C

Dieser Abschni dieser Anleitung wird nur dann benö gt, wenn Sie den Kalibrator mit eigener PC-
So ware ansteuern bzw. auslesen möchten. Diese Informa onen richten sich ausschließlich an
Fachpersonal mit entsprechenden Hard-, und So ware- sowie Programmier-Kenntnissen.
General characteris cs:
The communica on runs in half duplex way which means that the transmission and recep on of data cannot be
contemporeneously present. The regulator of the calibrator replies only a er receibing a command, it never replies
itself. The command and reply are ASCII character strings, as detailed forward. The communica on program will be
able to convert ASCII to decimal to extract numeric values. The default adress is 1. Baud rate: 2400, 4800, 9600 and
1200 baud, the default value is 9600; the other parameters are standard and cannot be changed.
Variables
0
Set point
1
Ramp ON/OFF
2
Set point 2
3
Gradient
4
Resolu on
5
Prop. Band
6
Integral me
7
Deriva ve me
8
Sensor input selec on
9
Title
10
Units
13
Access key
14
Baud rate
15
Adress
16
Serial number
18
Max. set point
19
Min. set point
21
Wait ON/OFF
22
Switch on temperature
23
Switch off temperature
24
Version
25
EXTernal sensor type
26
REFerence sensor type
28
Stability range
29
Symbol of steadiness
100 Temperature (internal sensor)
105 Temperature (EXTernal sensor)
106 Temperature (REFerence sensor)
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • DT-Info@Leitenberger.de •
Baud Rate:
Data bits:
1 = internal probe
2 = internal probe + EXTernal probe
3 = internal probe + REFerence probe
4 = internal probe + EXT + REF probe
0 = °C
1 = °F
2 = K
0 = Pt 100 (4-wire)
1 = thermocouple type N
2 = thermocouple type K
3 = thermocouple type J
4 = thermocouple type R
5 = thermocouple type S
6 = Pt 100 (3-wire)
7 = thermocouple type E
DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
(nur in Englischer Sprache verfügbar)
9600
Parity:
8
Stop bits:
Read / Write
38
No
1
Type
read / write
float
read / write
integer
read / write
float
read / write
float
read / write
integer
read / write
float
read / write
float
read / write
float
read / write
integer
read / write
float
read / write
integer
read / write
float
read
float
read / write
float
read
float
read
float
read
float
read
float
read
float
read
float
read
float
read / write
integer
read / write
integer
read
float
read
float
read
float
read
float
read
float
www.LR-Cal.net