1. Plug the adapter cable into the mains adapter socket
the blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
11. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description
Potential cause and remedy
«ERR 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
the measurement.*
«ERR 2»
Error signal
During the measurement, error signals
were detected by the cuff, caused for
27
instance by movement or muscle
tension.Repeat the measurement,
keeping your arm still.
aponorm
Professionell
®
Error
by microlife
®
«ERR 3»
24
«ERR 5»
in
4
«ERR 5»
«HI»
«LO»
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
Description
Potential cause and remedy
No pressure
An adequate pressure cannot be gene-
in the cuff
rated in the cuff. A leak may have
occurred. Check that the cuff is cor-
rectly connected and is not too loose.
Replace the batteries if necessary.
Repeat the measurement.
The measuring signals are inaccurate
Abnormal
and no result can therefore be dis-
result
played. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
AFIB/MAM-
There were too many errors during the
Mode
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 300 mmHg) OR the pulse is too
high
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
The pulse is too low (less than 40
Pulse too
beats per minute). Repeat the
low
measurement.*