Diagnostic messages
Error code
Meaning
E06
Saving the sen-
sor settings has
failed
E07
Invalid model
E08
Short-circuit on
digital outputs,
no output of test
results
E09
Load voltage too
low
E10
Invalid trigger
E11
Internal error
24
Operating Instructions, 10/2005 Edition, J31069-D1287-U001-A5-7418
Possible causes Corrective actions
Fault on saving
A taught-in
model was not
found for the
selected model
number with the
operating mode
set.
Short-circuit on
digital outputs
Overload on digi-
tal outputs, sup-
ply voltage too
low
Triggering when
sensor is not
ready
Temporary fail-
ure/fault; EMC;
heat generation,
overheating
Repeat calibration;
repeat teach-in
Use the correct
module number;
change the operating
mode
Eliminate short circuit
Check supply voltage;
rectify undervoltage
Note cycle times;
ensure triggering is
correct (max. every
33 ms in 200 ms)
Check the ambient
conditions, switch off
the internal lighting
SIMATIC MV220