Important:
The soldering tool becomes hot during the calibration
process. Never leave combustible materials near the hot
soldering iron
Calibration of the control unit (without soldering iron) can also
be carried out by a calibration laboratory. The calibration
values for the soldering iron are simulated.
4.2.2 Change Calibration
Press the DOWN key
Calibration at 100°C/212°F
The station adjusts the temperature of the soldering pencil to
100°C/212°F. Once the temperature becomes static (at
which point the indicator flashes), the soldering tip tempera-
ture (external measuring device) is compared to that shown
on the display. If a temperature deviation is found, the UP/
DOWN keys can be used to make adjustments. A maximum
temperature adjustment of ±40°C/±72°F is possible. If the
measured temperature matches that shown on the display,
press the II key (CAL) to confirm by temperature deviation is
reset to 0. This concludes the calibration at 100°C/212°F.
Press the III key (EXIT) to exit the menu without saving any
changes.
Temperature
deviation
Press the UP key
Calibration at 450°C/842°F
The station adjusts the temperature to 450°C/842°F. Once
the temperature becomes static (at which point the indicator
flashes), the soldering tip temperature (external measuring
device) is compared to the actual value shown on the display.
If a temperature deviation is found, the UP / DOWN keys can
be used to make adjustments. A maximum temperature
adjustment of ±40°C/±72°F is possible. If the measured
temperature matches that shown on the display, press the II
key (CAL) to confirm by the temperature deviation is reset to
0.
This concludes the calibration at 450°C/842°F.
Press the III key (EXIT) to exit the menu without saving any
changes.
39
CAL
EXIT
After both calibration points, 100°C (212°F) and 450°C
(842°F), have been calibrated and confirmed, the calibration
process is complete.
Temperature
deviation
4.2.3 Station code (ID number)
When using multiple WD stations, you can assign a number
from 0 - 999 to each soldering station for identification pur-
poses.
If the optional USB interface is used, several WD stations,
with the full range of functions, can be remote-controlled. For
this purpose, each WD station requires a station code for
unambiguous identification (ID number).
Use the UP / DOWN buttons to change the ID number.
Switch to previous menu item with I.
Switch to next menu item with II.
Button III is used to leave the menu 2 again (EXIT).

5. USB Interface

The WD 1M control unit is equipped with a Mini USB interface
as standard. This can retrofitted on the WD 1 control unit.
(See Page 167). WELLER standard software (CD is included in
scope of supply) has been provided to use the USB interface).
This software contains a firmware updater and monitor soft-
ware.
The firmware updater can be used to carry out a software