Herunterladen Inhalt Inhalt Diese Seite drucken

Filter Replacement; Self-Diagnostic And Error Messages; Replacement Parts / Accessories - E+E Elektronik EE375 series Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.2

Filter replacement

A soiled filter should not be cleaned but replaced with a new one. The order number for
new filters is HA010103.
6.3

Self-diagnostic and error messages

i
Definition:
• Error:
- possible cause
=> measures / help
• Long response time:
- Filter soiled
=> Replace filter
• Transmitter failure:
- No supply voltage
=> Check wiring and supply voltage
=> contact the manufacturer
• High humidity values
- Dew (condensation) in the sensor probe head
=> heat the measurement cell and check the mounting of sensor probe
7.

REPLACEMENT PARTS / ACCESSORIES

- Sampling cell with quick connector
- Stainless steel sintered filter
- Basic sampling cell
- Configuration software + interface cable
When replacing filters, please note the following:
- Unscrew the filter cap very carefully to avoid
damaging the sensing element.
- Handling the filter might clog the pores.
Use gloves to screw on the new filter.
(HA050102)
(HA010103)
(HA050103)
14

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis