3. The output data format
Changing according to the initial settings (See P16.).
q Normal condition
Initial setting :
Normal output (set at time of shipping)
1st byte
2nd byte
3rd to 9th bytes : Numerical data (ex.12.3456)
Initial setting :
1st byte
2nd byte
3rd byte
4th byte
5th to 11th bytes : Numerical data (ex. 00.0000)
40
(E)
: Channel name (A or B)
: Sign ("+" or "–")
(Outputting with measurement
mode information)
: Channel name (A or B)
: Current mode
( N : Current value,
P : Peak-to-peak value,
I : Minimum value,
A : Maximum value)
: Unit (M: mm, I: inch)
: Sign ("+" or "–")
Initial setting :
Outputting with measuring mode information
and comparator Go/No Go result
1st byte
2nd byte
3rd byte
4th byte
5th byte
6th to 12th bytes : Numerical data
: ("+" or space)
: Channel name (A or B)
: Current mode
(N: Current value,
P: Peak-to-peak value,
I : Minimum value,
A: Maximum value)
: Unit (M: mm, I: inch)
: Comparator Go/No Go
result
U : Upper limit over
G : Within range
L : Lower limit under
E : When an alarm has
occurred
: Sign ("+" or "–")
(ex.00.0000)