5. Commissioning, operation
5.6 Zero adjustment (only PG43SA-S)
External zero adjustment, setting range ±15°
5.7 Autoclaving
The following applies only to model PG43SA-S with option "autoclav-
able version", recognizable on the dial printing as "AUTOCLAVABLE".
Observe the permissible conditions of autoclaving, see chapter
9.3 "Permissible temperature"
During the autoclaving process, no weight-loading must be exert-
ed via the case and the instrument must not lie on its window.
After the first autoclaving cycles, the window slightly arches
outward. This has no effect on the function of the instrument.
Depending on the autoclaving process, whitish residues can build
up on the window and pointer. This has no effect on the function of
the instrument.
In case of a zero offset, carry out a zero adjustment. See chapter
5.6 "Zero adjustment".
WIKA operating instructions pressure gauge, PG43SA-S, PG43SA-C
The zero adjustment can be
made with a slotted screwdriver
after the removal of the sealing
plug on the top of the case.