6. Commissioning, operation
6.5.1.2 Setting test item 2
The test item 2 is programmed in the same way as test item 1.
If only 1 test item is measured, the type of test item [---] is selected in the selection window of test items 2 and 3.
[Next ▶] allows further logger parameters to be defined.
6.5.1.3 Setting test item 3
1. Press button [Start ▶].
⇒
The programming window for test item 3 is displayed.
2. Press button
1
.
⇒
The selection window of the possible types of test item is displayed.
⇒
Selection = "DIFF" always calculates test item 2 - test item 1
3. Select types of test item.
⇒
Return to the menu screen.
6.5.1.4 Setting logger parameters
Under the setting of "Test item 3", the logger parameters are also set.
By selecting "Automatic" the following parameters are entered:
Automatic logger function
■
Manual logger function
■
8
⇒
Data points are recorded by manual confirmation.
Interval adjustable
and
■
9
Duration adjustable
and
■
11
Start time
and
: current or freely selectable start date
13
14
■
⇒
Observe the system time!
Setting the interval
9
1. Press button
.
10
⇒
A numeric keypad will be displayed.
2. Enter the number of intervals.
3. Confirmed with [OK].
⇒
Return to the menu screen.
Setting the duration
11
4. Press button
12
.
⇒
A numeric keypad will be displayed.
5. Enter the duration.
6. Confirmed with [OK].
⇒
Return to the menu screen.
Setting the start time
13
7. Press button
14
.
⇒
A numeric keypad will be displayed.
8. Enter a desired time or the current time.
9. Confirmed with [OK].
⇒
Return to the menu screen.
10. Press [Start ▶] to start the logger process.
WIKA operating instructions portable process calibrator, model CPH7000
7
: 0.5 ... 3,600 s
10
: 1 s ... 60 h
12
7
8
9
10
11
12
13
14
45
EN