Care
Contamination of the apparatus, especially the admission pressure
connection, must be avoided.
Attention: only use in clean condition!
After a measurement in which the contamination levels are higher than
the standards allow, scavenge the apparatus with clean air for at least
6 minutes.
Dismantle detection unit during greater impurity and clean compo-
nents with soap water.
After this dry with oil-free and dry air.
While assembling ensure, that the feeders are screwed in correct prec-
edence.
8
Inspection
The measuring unit should be recalibrated if incorrect measurements
are suspected.
Inspection and servicing should be carried out annually
9
Fault – Cause – Remedy
.
Fault
Cause
Snap-on nipple
Sealing ring worn
connection leaking
Tube holder faulty Material wear
Measurement er-
Delivery unit
ror
blocked
SimultanTest CO2
Remedy
Use new sealing ring;
replace sealing ring
Replace tube holder and clip
Bubble check, flow quantity
check with nozzle and hose
in water glass.
10

Technical Data

Medium:
Standard operating pres-
sure:
Flow:
Weight (including case):
Dimensions (case):
height / width / length
Measuring limits
Carbon dioxide CO
2
3 bar ±0.2
The measuring unit is set to the flow
necessary for the Dräger-Tubes
0.2 L/min NH
, NO
, CO, SO
3
x
2
0.8 L/min H
S and
2
2.0 L/min H
O.
2
approx. 3 kg
91 x 300 x 400 mm
See Measuring range on page 10 or In-
struction for Use of the Dräger-Tubes
Care
®
of
, PH
,
3
®
13