Recommended separation distances between
portable and mobile RF communications equipment and the MicroCO
The MicroCO is intended for use in an e lectromagnetic environment in which
radiated R F disturbance s are controlled. The customer or the user of the
MicroC O can help prevent electromagnetic interference by maintaining a
m inimum distance betwe en portable and mobile RF communications equipment
(transmitters) and the MicroCO as re commended below, according to the
m ax imum output powe r of the communications equipment.
Rated Maximum
Output Power
of
Transmitter in
Watts (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output powe r not listed above, the
re com mended separation distance d in m eters (m) can be e stimated using the
e quation applicable to the fre quency of the transmitter, whe re P is the
m ax imum output powe r rating of the transmitter in watts (W) according to the
transmitter m anufacturer.
NO TE 1
At 80 MHz and 800 MHz, the sep aration distance for the higher
fre quency range applies.
NO TE 2
The se guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and re flection from structures obje cts and
pe ople
Separation Distance in Meters (m) according to
Frequency of Transmitter
150 KHz to
80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
80 MHz to
800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
69
800 MHz to
2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23.3