Herunterladen Inhalt Inhalt Diese Seite drucken

CareFusion MicroCO Meter Betriebshandbuch Seite 67

Vorschau ausblenden Andere Handbücher für MicroCO Meter:
Inhaltsverzeichnis

Werbung

WARNING:
The MicroCO should not be used adjacent to or stack ed with other
e quipment. If adjacent or stack ed use with other e quipm e nt is ne ce ssary, the
MicroC O and the other e quipme nt should be obse rve d /m onitore d, to ve rify
norm al ope ration in the configuration in which it will be use d.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The MicroCO is intended for use in the electromagnetic environment specified
be low. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Emission Test
R F e m issions
C ISPR 11
R F e m issions
C ISPR 11
Harm onic emissions
IEC 61000-3-2
Voltage fluctuations /
flick e r e missions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroCO is intended for use in the electromagnetic environment specified
be low. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Immunity
Test
Ele ctrostatic
discharge
(ESD)
IEC 61000-4-2
Ele ctrical fast
transient /
Compliance
Group 1
Group B
Not Applicable
(batte ry
powe re d)
Not Applicable
(batte ry
powe re d)
IEC 60601 test
level
± 6 k V contact
± 8 k V air
± 2 k V for
powe r supply
Electromagnetic Environment -
Guidance
The MicroCO uses RF energy only
for its inte rnal function. Therefore ,
its R F e missions are very low and
are not likely to cause any
inte rfe rence in nearby e lectronic
e quipment.
The MicroCO is suitable for use in all
e stablishments, including domestic
e stablishments and those directly
conne cted to the public low-voltage
powe r supply network that supplies
buildings used for domestic
purposes
Compliance
Level
± 6 k V
contact
± 8 k V air
Not Applicable
(Batte ry
65
Electromagnetic
Environment -
Guidance
Floors should be wood,
concre te or ceramic tile.
If floors are covered
with synthetic m aterial,
the re lative humidity
should be at least 30%.
Mains powe r quality
should be that of a

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis