Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroCO is intended for use in the electromagnetic environment specified
be low. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Immunity
Test
C onducted
R F
IEC 61000-4-
6
R adiated R F
IEC 61000-4-
3
IEC 60601
Compliance
test level
Level
3 Vrm s
150 k Hz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
Electromagnetic Environment -
Guidance
Portable and m obile R F
com m unications equipment
should be used no closer to any
part of the MicroCO, including
any cables, than the
re com mended separation
distance calculated from the
e quation applicable to the
fre quency of the transmitter.
Recommended separation
distance (d)
3 Vrm s
d = 1.2√P
3 V/m
d = 1.2√P
800 MHz
d = 2.3√P
2.5 GHz
W he re P is the maximum output
powe r rating of the transmitter
in watts (W ) according to the
transmitter m anufacturer and d
is the re commended separation
distance in meters (m).
Fie lds stre ngths from fixed RF
transmitters, as determined by
an e le ctromagnetic site survey,
should be less than the
com pliance level in each
fre quency range.
Inte rference may occur in the
vicinity of e quipment marked
with the following s ymbol:
67
80 MHz to
800 MHz to
a
b