Herunterladen Inhalt Inhalt Diese Seite drucken

Adjust The Spring Tension To The Strength Required By The Patient - Albrecht Individual Cds Joint For Dynamic Redression Gebrauchsanweisung

Inhaltsverzeichnis
Instructions For Use
3.4. Adjust the spring tension to the strength required by
the patient
1
The set spring force is indicated by a scale from 0 to 15 and can be seen
through the viewing window on the CDS
below 0 are marked in red.
To prevent damage to the CDS
the red area of the CDS ® Nano hybrid joint viewing window .
2
Insert the tool into the hole on the side of the joint until it stops. Turning
it clockwise or in the + direction increases spring tension, turning it
counterclockwise or in the - direction decreases spring tension.
1 1
If two joints are used, the spring tension must be the same for both joints.
Activating or deactivating the spring tension does not change the strength of the
spring tension.
The spring force may only be adjusted in consultation with the attending
physician.
®
housing,. The areas above 15 and
Nano hybrid joint, do not rotate into
®
2
2
2 2
www.albrechtgmbh.com