2.3.1 Step 1
Stand in a natural position, then press the calibrate button in the mobile application, and remain still
until the prosthesis completes the calibration. Moving the prosthesis during this procedure may result in
improper functioning.
2.3.2 Step 2
Put the tip of the shoe against a wall, for example. Position the ankle joint in neutral, so that the upper
part of the prosthesis is perpendicular to the ground. Press the calibrate button, moving the prosthesis
during this procedure may cause it to its malfunction.
2.3.3 Step 3
In this step, perform dorsal flexion of the prosthesis (for example, by performing a squat), the exact
positioning of the prosthesis is shown in Figure 3. Then, press the calibrate button and, in stillness, wait
for the prosthesis to perform calibration. Moving the prosthesis during this procedure may cause it to
malfunction.
2.3.4 Step 4
Stand in a natural position, enter your weight in the app, then press the calibrate button. In stillness, wait
for the prosthesis to perform the calibration. Moving the prosthesis during this procedure (changing the
load on the prosthesis) may cause it to malfunction.
2.3.5 Step 5
In the next stage of calibration, you will need to go through several walking steps. When you are ready,
press the START button and walk a minimum of 5 steps. Once done, press the STOP button.
Figure 3: Prosthesis position in dorsal flexion.