SITRANS F M MAGFLO
Maximum allowable torques
(to meet PED)
Nominal
size
EN 1092-1
EN 1092-1
PN 6
PN 10
mm Inch Nm F/Lbs Nm
15
½" N/A
N/A
N/A
25
1"
N/A
N/A
N/A
40
1½" N/A
N/A
N/A
50
2"
N/A
N/A
N/A
65
2½" 10
7
N/A
80
3"
25
18
N/A
100
4"
25
18
N/A
125
5"
25
18
N/A
150
6"
25
18
N/A
200
8"
25
18
50
250 10"
25
18
50
300 12"
50
37
60
350 14"
50
37
60
400 16"
50
37
88
450 18"
56
41
92
500 20"
53
39
103
600 24"
81
60
161
Torque calculations
Manufacturer's design
and safety statement
Electromagnetic flowmeter type MAG 3100 with PTFE or PFA liner
EN 1092-1
EN 1092-1
PN 16
PN 25
F/Lbs Nm F/Lbs
Nm F/Lbs
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
25
18
N/A
N/A
N/A
25
18
N/A
N/A
N/A
25
18
N/A
N/A
N/A
32
24
N/A
N/A
N/A
50
37
N/A
N/A
37
52
38
105
77
37
88
65
160
118
44
117
86
170
125
44
120
89
240
177
65
170
125
330
244
68
170
125
320
236
76
230
170
390
288
119
350
258
560
413
All values are theoretical and are calculated making the following assumptions:
1.
All bolts are new and material selection is according to EN 1515-1 table 2.
2.
The standard flange gasket material to be used between the flowmeter and mating flanges
should not exceed 75 shore A durometer.
3.
All bolts are galvanized and adequately lubricated.
4.
The values are calculated for use with carbon steel flanges.
5.
Flowmeter and mating flanges are correctly aligned.
1.
Responsibility for the choice of lining and electrode materials with regard to their abrasion
and corrosion resistance lies with the purchaser; the effect of any change in process
medium during the operating life of the flowmeter should be taken into account. Incorrect
selection
of lining and/or electrode materials could lead to a failure of the flowmeter.
2.
It is the responsibility of the user to ensure that stresses and loading caused by earthquakes,
traffic, high winds and fire damage are taken into account during installation, when
appropriate. These forces are not taken into account during flowmeter design.
3.
It is the responsibility of the user to ensure that the flowmeter is installed such that it does
not act as a focus for pipeline stresses. External loadings are not taken into account during
flowmeter design.
4.
During operation do not exceed the pressure and/or temperature ratings indicated on the
data label or in the installation instructions.
5.
It is the responsibility of the user to ensure that all installations include over pressure
protection, means for draining/venting, and that adequate protection is provided to
minimise any risk of contact with hot surfaces.
6.
Under the Pressure Equipment Directive this product is a pressure accessory, and not
approved for use as a safety accessory, as defined by the Pressure Equipment Directive.
7.
Removal of the terminal box except by Siemens Flow Instruments or their approved agents
will invalidate the PED conformity of the product.
In accordance with the Pressure Equipment Directive (97/23/EC)
Maximum torque
EN 1092-1 ANSI B16.5 ANSI B16.5 AS 2129
PN 40
Class 150 Class 300
Nm F/Lbs
Nm F/Lbs
Nm F/Lbs Nm F/Lbs Nm F/Lbs Nm F/Lbs
10
7
N/A
N/A
3
16
12
7
5
7
34
25
9
7
18
46
34
20
14
11
34
25
25
18
16
42
31
34
25
24
72
53
26
19
34
114
84
42
31
42
144
106
57
42
35
185
137
88
65
57
300
221
99
73
57
320
236
132
97
86
450
332
225
166
N/A
650
480
210
155
N/A
570
421
220
162
N/A
740
546
200
148
N/A
1220 900
280
207
N/A
JIS
Table E
K10
2
6
4
8
6
5
9
7
16
12
13
17
13
20
15
8
20
15
27
20
12
21
16
36
27
18
32
23
21
16
25
24
18
26
19
31
33
24
43
32
26
40
30
76
56
42
61
45
46
34
42
62
46
72
53
63
78
58
56
41
N/A
147
109
72
53
N/A
168
124
115
85
N/A
164
121
109
80
N/A
194
143
127
93
N/A
306
225
150
111
JIS
K20
8
6
16
12
20
15
13
10
18
13
28
21
37
27
59
44
40
30
62
45
113
83
96
71
147
108
175
129
184
135
211
155
255
188
7